Monochromated X-Ray Source with XR5 Electron Gun 

Issue 5 

 

Monochromator and XR5 (HA600054)

13

5.  Note the position of the beam spot relative to the cross hairs centre. 

6.  If the spots are less than 1 mm from the cross hairs centre the spot can 

probably be re-centred without causing a significant change in the detection 
sensitivity.

7.  Record a spectrum from any convenient sample (e.g. a silver standard) before 

making any changes and then use the two lateral shift micrometers on the 
crystal manipulator as shown in to adjust the beam position to the cross hairs 
centre. Do not adjust the height or leg angle of the crystal manipulator. Re-
record the spectrum as before and if the count rate has not fallen by more than 
5% there is no need to realign further. If the spot is more than 1 mm from the 
centre or if this procedure needs to be repeated on more than three occasions, 
then a more complete realignment should be performed. Only if a major service 
or repair has been undertaken should a more extensive re-alignment be 
necessary.

Figure 3-6: Adjustment of the Crystal Manipulators 

Lateral Shift 
Micrometers

Lateral Shift 
Micrometers

Height Change = 
Diagonal Movement of 
Spot (

NOT

 Wanted!) 

Dispersive Change = 
Vertical Movement of Spot
(Along HPT X-axis) 

Non-Dispersive Change = 
Horizontal Movement of Spot 
(On the HPT Y-axis) 

Monochromated X-Ray Source with XR5 Electron Gun

Monochromator and XR5 (HA600054) 

 

Issue 5  

14

3.7 

T352/NT Computer-Controlled Power Supply 

The power supply is a multiple output unit generating all the voltages needed to 
operate both the Twin Anode and Monochromator guns. It also controls the cooling 
water pumping for both guns. 

The specification of the outputs for the monochromator gun are shown below in an 
abbreviated form. 

Output

Maximum Operating Ratings

Main HT

15kV/60mA (Normal operating) 
or 20kV/20mA (Source conditioning)

Monochromator Filament

2.5A

Bombard

3000V

Cathode

1200V

Focus

3500V

Table 2: Specification Outputs for the Monochromator Gun 

Control is entirely from the computer, no manual controls are fitted except for the rear 
panel mains power switch. 

The supply requires five interlock loops: Vacuum; Water-flow; Twin anode cover; Mono 
cover; Mono source cable-guard, to be closed before the high voltage outputs can be 
switched on. The state of the interlocks is reported to the software. To clear an 
interlock the physical fault must first be rectified (e.g. replace a safety cover) and then 
reset by software before the high voltage outputs can be switched on.  The interlocks 
require a voltage free closing contact to operate. Shorting interlocks to earth will not 
operate them. 

Communication between the control computer and the power supply is via an optical 
modem and a fibre optic serial line from an RS232 port (generally COM1). The mains 
power to the T352/NT is NOT vacuum-interlocked (although the outputs are) so that 
the communications circuitry is always powered even when T352/NT output trips 
occur. The output circuitry is protected by the separate vacuum interlock. 

The T352/NT outputs will trip if any interlock fails or if certain levels are exceeded. The 
most relevant of these are the Leakage Current and the HT Current. The leakage 
current is due to the conductivity of the anodes cooling water and the HT current is the 
total current comprising leakage c electron-beam current. Trips will occur if the 
leakage current exceeds 20mA, (so water purity must be maintained periodically if the 
drain current becomes too high), or if the HT current exceeds 60mA. 

The state of all outputs, interlocks, trips and the values of achieved voltages and 
currents are displayed by the control software. 

There are no user serviceable parts in the T352/NT power supply. Contact your local 
agent for service. 

Summary of Contents for ESCALAB 250

Page 1: ...Moving the XR5 Anode 12 3 6 Monochromator Fine Tuning 12 3 6 1 Alignment of the Analysis Centre and Microscope Cross Hairs 12 3 7 T352 NT Computer Controlled Power Supply 14 4 Monochromator Operation 15 4 1 Avantage Data System Monochromator Controls 15 4 1 1 Errors and Warnings 16 4 2 Different XR5 Operating Modes 16 4 3 Switching on the Re circulated Water 17 4 4 Observing the Monochromator Spot...

Page 2: ...ing the online manual loaded with the software datasystem It does not include detail on the initial installation and setting up of the electron gun or crystals nor any troubleshooting or maintenance which would require the user to directly access high voltage It includes procedures for replacement of certain consumable items such as anodes and filaments which require the system to be vented and su...

Page 3: ...nterlocks are displayed in the software within the Trips Interlocks display as in GREEN OK or RED FAIL for the state of the interlocks Typical Trips Interlocks displays are shown in Figure 2 1 Occasionally pop up warning messages will appear After taking note of the message select the OK button Go into the details window and look at the Trips Interlocks you will see a red FAIL in the appropriate w...

Page 4: ... software they include x Temperature in T352 Power Supply rises to unacceptable levels x Current drawn from T352 Power Supply HV greater than 60 mA x Drain from HV through X ray water lines greater than 20 mA Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA600054 Issue 5 8 3 Description 3 1 Introduction The Monochromated X Ray Source produces X Rays which have well defined...

Page 5: ...uit uses mains water or a re circulator as its input and then increases the flow through the X Ray source by use of a booster pump The control of the booster pump and of the solenoid valves which switch the water flow between the twin anode if fitted and mono source is by the datasystem via the T352 NT power supply Both the twin anode and mono anode are floating at high voltage and have water flow...

Page 6: ...e by 0 8 mm per turn The range of motion and the useful anode length are both 25 mm A fraction of a turn for example a half turn will usually be enough to expose a new aluminium film area 3 6 Monochromator Fine Tuning Complete alignment and set up of standard operating conditions will generally be carried out by Thermo Fisher Scientific service engineers and most operators will not need to be conc...

Page 7: ...ing Ratings Main HT 15kV 60mA Normal operating or 20kV 20mA Source conditioning Monochromator Filament 2 5A Bombard 3000V Cathode 1200V Focus 3500V Table 2 Specification Outputs for the Monochromator Gun Control is entirely from the computer no manual controls are fitted except for the rear panel mains power switch The supply requires five interlock loops Vacuum Water flow Twin anode cover Mono co...

Page 8: ...the Apply button should then be selected Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA600054 Issue 5 16 4 1 1 Errors and Warnings As the monochromator is ramping up to the desired operating point the data system will check for any problems and report these as a warning window If an Error does occur the data system will display a warning dialogue and instruct the operato...

Page 9: ...upplied from the system or from a separate mains circuit in the laboratory If this is not switched on then when an attempt is made to switch on the monochromator the booster pump will run dry be much louder than normal and possibly be damaged The monochromator gun will not turn on as the water flow interlock will have failed 4 4 Observing the Monochromator Spot For fine tuning of the monochromator...

Page 10: ...nd converge as the sample is raised and lowered 4 Position each spot by moving the dispersive and non dispersive adjusters on the crystal manipulator do NOT adjust the heights of the crystal as this will change the energy of the X Ray spot from each crystal to coincide with the cross hairs on the viewing camera These adjusters are very sensitive so only small adjustments should be made Never adjus...

Page 11: ...eplacing Anodes In general the anode will need replacing only when all the aluminium coating has been degraded sufficiently to attenuate the X Ray flux below acceptable levels along its entire length If the anode condition requires its replacement the vacuum system will have to be fully shut down and vented to dry nitrogen Figure 5 3 XR5 Anode Assembly Refer to Figure 5 3 XR5 Anode Assembly 1 Unpl...

Page 12: ... vented to dry nitrogen As a further precaution disconnect the mains supply from the T352 NT X Ray power supply Refer to the Figure 5 5 XR5 Lens and Emitter Assemblies 1 Remove the M8 x 16 hex head bolt clamping the XR5 earth strap 2 Unplug the leads from the XR5 3 Do not unbolt the 70mm FC38 11 pin feedthrough flange This is unnecessary except when replacing the feedthrough 4 Remove all the M8 x ...

Page 13: ...ges Use a new FC64 copper gasket and make sure that the flange is offered to the gun in the correct orientation The main flange is labelled TOP on its circumference adjacent to the focus connection This should be in the top vertical position when refitted to the gun on the system 22 Check the pin base for filament and interlock continuity and all other pins for electrical isolation 23 Fully tighte...

Page 14: ...Issue 5 28 5 9 Fault Finding Symptom General Cause Specific Cause How to Rectify How to Fix Mains re circulated water supply switched off Booster pump will make excessive noise Check supply Look for warning messages on datasystem Switch on supply Insufficient water flow Warning message on datasystem Water flow not sufficient to satisfy flow requirement Check inlet and outlet flow Look for warning ...

Page 15: ...n still runs correctly Loss of control of the gun from the datasystem Gun still running An HT tick has effected communication Reboot the datasystem Try the gun control again Table 4 Fault Finding 5 10 Fuses Information for Service Engineers Within the digital X ray PSU T352 NT there is an 800mA T fuse on the mains input Monochromated X Ray Source with XR5 Electron Gun Monochromator and XR5 HA60005...

Reviews: