Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
73
a.
Remove the digital board, as described in "
Replacing the Digital Board" on page 87
b. Disconnect cables from their connectors.
c.
Remove the six nuts holding the analog board in place.
d. Remove the analog board.
5. Install the new analog board (PN 119590-0022).
a.
Insert the new analog board.
b. Reattach the six nuts.
c.
Reconnect the cables.
d. Reattach the digital board.
6. Reattach the left cover of the DSQ II.
7. Restart the DSQ II.
Replacing the Analog Board Fuse
To replace the analog board fuse:
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Replace the analog board fuse.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug
the DSQ II to avoid an electrical shock.