Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
87
Replacing the Digital Board
To replace the digital board, you will need:
• New Digital board (PN 119590-0010)
• Nut driver, 5.5 mm
• Wrench, Allen, 2.5 mm
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Remove the digital board.
Figure 48.
Removing the Digital Board
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Two Nuts Around the Digital
Board Cover
5
Six Nuts Around the Digital Board
2
Connector to Ethernet Cable (to
PC)
6
Connector to J3 Cable (to
Electrometer Board)
3
Seven Screws Around the Digital
Board Cover
7
Four Screws Around the Digital
Board Cover
4
Digital Board Cover
8
Digital Board (PN 119590-0010)
7
5
6
8
1
2
3
7
1
4
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...