Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
103
Replacing the Power Module Fuses
To replace a power module fuse, you will need:
• Power Module Fuses F1, F2 (PN A0006-04550; 0.315 A, 5 x 20 mm, 250 V)
• Power Module Fuse F3 (PN A0006-06075; 800 mA, 5 x 20 mm, 250 V)
• Power Module In-line Fuses (PN A0006-10510; 3.15 A, SB, 5 x 20 mm, 250 V)
• Screwdriver, Phillips #2
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the top and front covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the current power module.
Figure 61.
Removing the Power Module
a.
Remove the two screws holding the power module in place.
b. Lift the power module straight up.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Power Module
3
Power Module Cover
2
Two Screws Around the Power Module
3
1
2
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...