Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
86
DSQ II Hardware Manual
Thermo Scientific
a.
Remove the lower screw and only loosen the upper screw holding the DC driver
cover in place.
b. Slide the cover up off the upper screw to remove it.
c.
Remove the DC driver cover.
5. Remove the current DC driver fuse.
6. Replace the DC driver fuse (PN 76703-0125; 1.25 A, SB, 5 x 20 mm, 250 V).
Figure 47.
Replacing the DC Driver Fuse
7. Reattach the DC driver cover and reattach the screw that holds the cover in place.
8. Reattach the left, top, and front covers of the DSQ II.
9. Restart the DSQ II.
DC Driver Fuse
CAUTION - INSTRUMENT DAMAGE.
Only use a 125V fuse or else you could damage the
instrument. Never replace DSQ II fuses with those of a different type, voltage, or current
rating. Also, to avoid damaging the boards, you should always wear an electrostatic
discharge (ESD) strap on your wrist when replacing fuses.