8 • EG Troubleshooting
Doc. 065342-06 12/11
255
8.23
No Flow
•
DP/SP power is off
Turning off the DP/SP automatically turns off the EG and the suppressor.
Current to the CR-TC is automatically turned on and off when the EG
power is turned on and off.
Check that the power to the DP/SP is turned on. Prime the pump (see
) and resume operation.
•
DP/SP pressure limit tripped
Verify that the
Current Pressure
reading on the Chromeleon 7 ePanel Set
or the Chromeleon 6.8 panel tabset is between the high and low limits
displayed on the panel. For details, see
NOTE
The analytical EGC requires at least 14 MPa
(2000 psi) of backpressure for optimal removal of
electrolysis gas from the eluent produced by the car-
tridge. A system backpressure of 16 MPa (2300 psi) is
ideal.
•
(Analytical IC only) RFIC Eluent Degasser tubing is ruptured
If flow from the EG waste line is normal but there is no flow through the
columns, the tubing assembly inside the analytical IC system RFIC Eluent
Degasser has ruptured. Replace the degasser (see
Summary of Contents for Dionex ICS-5000
Page 66: ...Dionex ICS 5000 Operator s Manual 50 Doc 065342 06 12 11 ...
Page 130: ...Dionex ICS 5000 Operator s Manual 114 Doc 065342 06 12 11 ...
Page 168: ...Dionex ICS 5000 Operator s Manual 152 Doc 065342 06 12 11 ...
Page 176: ...Dionex ICS 5000 Operator s Manual 160 Doc 065342 06 12 11 ...
Page 184: ...Dionex ICS 5000 Operator s Manual 168 Doc 065342 06 12 11 ...
Page 200: ...Dionex ICS 5000 Operator s Manual 184 Doc 065342 06 12 11 ...
Page 206: ...Dionex ICS 5000 Operator s Manual 190 Doc 065342 06 12 11 ...
Page 210: ...Dionex ICS 5000 Operator s Manual 194 Doc 065342 06 12 11 ...
Page 214: ...Dionex ICS 5000 Operator s Manual 198 Doc 065342 06 12 11 ...
Page 256: ...Dionex ICS 5000 Operator s Manual 240 Doc 065342 06 12 11 ...
Page 274: ...Dionex ICS 5000 Operator s Manual 258 Doc 065342 06 12 11 ...
Page 318: ...Dionex ICS 5000 Operator s Manual 302 Doc 065342 06 12 11 ...
Page 348: ...Dionex ICS 5000 Operator s Manual 332 Doc 065342 06 12 11 ...
Page 374: ...Dionex ICS 5000 Operator s Manual 358 Doc 065342 06 12 11 ...
Page 422: ...Dionex ICS 5000 Operator s Manual 406 Doc 065342 06 12 11 ...
Page 434: ...Dionex ICS 5000 Operator s Manual 418 Doc 065342 06 12 11 ...
Page 438: ...Dionex ICS 5000 Operator s Manual 422 Doc 065342 06 12 11 ...
Page 448: ...Dionex ICS 5000 Operator s Manual 432 Doc 065342 06 12 11 ...
Page 454: ...Dionex ICS 5000 Operator s Manual 438 Doc 065342 06 12 11 ...
Page 462: ...Dionex ICS 5000 Operator s Manual 446 Doc 065342 06 12 11 ...
Page 482: ...Dionex ICS 5000 Operator s Manual Index 20 Doc 065342 06 12 11 ...