4 • System Plumbing
Doc. 065447-02 12/14
93
2. Connect the Suppressor Bypass
REGEN IN
line to the ED cell
OUT
port.
3. Connect the waste line from the CR-TC
REGEN OUT
port to the gas
separator waste tube (see
4. Route the IC Cube tubing as described in
.
The flow path is ready for equilibration (see
5. If you are using a PdH reference electrode, follow the instructions in the
next section to condition the electrode before equilibrating the system.
Conditioning the PdH Reference Electrode
Condition the PdH reference electrode after plumbing the cell.
1. On the Chromeleon 7 ePanel Set or Chromeleon 6.8 panel tabset, select
the following settings:
a.
Pump panel: Set the pump flow rate to 0.01 mL/min. This turns on the
pump flow, also.
b. Eluent Generator panel: Enter 100 mM in the
Target Concentration
field. This turns on the power to the Dionex EGC, also.
c.
Verify that eluent is exiting the cell.
d. ED
panel: Set the reference electrode mode to
PdH
.
e.
Select DC amperometry mode, set the cell voltage to 0.000 V and
turn on the cell voltage. Switch to integrated amperometry mode and
select the PdH-referenced carbohydrate waveform.
2. Condition the electrode for 1 hour at these settings.
3. Turn off the pump flow. This turns off the Dionex EGC power and the cell
voltage.
4.1.14 Routing the IC Cube Tubing
1. For a neater appearance, and to ensure that the DC doors close properly,
route tubing that connects to other modules through the appropriate side
slots on the DC and the other modules.