Doc. 031857-03 4/06
F-1
F • FAQ
F.1
How do I hook up an autosampler?
For instructions on how to connect the ICS-2000 to an AS or AS50 autosampler,
see
of this manual. To connect an AS40, see
. Also refer
to the autosampler operator’s manuals, which are included on the Dionex
Reference Library CD-ROM (P/N 053891).
F.2
How do I print?
Click the Print toolbar button in Chromeleon (see
F.3
Why are the retention times moving?
Retention times can shift if the pump flow is erratic or if the column or eluent is
contaminated. See
for pump flow rate troubleshooting. If a
contaminated column is suspected, clean the column as instructed in the column
manual. The column manual is included on the Dionex Reference Library CD-
ROM (P/N 053891).
F.4
How do I adjust retention times?
Retention times are calculated during calibration. The
Use Recently Detected
Retention Time
parameter in the Chromeleon QNT Editor (
General
tab) can be
used to compensate for some types of retention time drifts; for example,
evaporation of volatile components in pre-mixed solvents or an aging column.
Refer to the Chromeleon online Help or user’s guide for details.
F.5
When should I remake standards?
Standards are used only for calibration and should always be made fresh (they
have a lifetime of only one week).
Summary of Contents for Dionex ICS-2000
Page 1: ...ICS 2000 ION CHROMATOGRAPHY SYSTEM OPERATOR S MANUAL Document No 031857 Revision 03 April 2006...
Page 64: ...ICS 2000 Ion Chromatography System 2 44 Doc 031857 03 4 06...
Page 204: ...ICS 2000 Ion Chromatography System A 6 Doc 031857 03 4 06...
Page 366: ...ICS 2000 Ion Chromatography System D 12 Doc 031857 03 4 06...
Page 372: ...ICS 2000 Ion Chromatography System F 4 Doc 031857 03 4 06...
Page 378: ...ICS 2000 Ion Chromatography System G 6 Doc 031857 03 4 06...