Maintenance
92
|
Dionex IC Pure system
Thermo Scientific
Chapter 13
Replacing the 0.2
μ
m final filter
7
Insert the ultrapure cartridge back into the unit (see “Changing the
ultrapure cartridge” on page 79). If an optional pretreatment car-
tridge is used, install it also in the unit (see “Replacing the pretrea-
ment cartridge (optional)” on page 80).
8
Switch the system on. If the unit is connected to an external 20 L
tank with level control, open the supply of feedwater.
9
c.
After entering the code push the “Menu and UV“ button
simultaneously. The display shows “UV Menu“.
d.
Push the “Menu“ button repeatedly until new UV-lamp
appears and press enter to confirm. The Display shows then
“Please wait“.
Step
Action
Figure
Note
For this action you must have a Level 3 code. Refer to
section “Entering a Code number” on page 71 to read where to
find this code and how to enter it.
Note
The system sets the operating hours counter of the
UV-lamp back and saves the new values by an automatically
calibration.
a.)
b.)
Step
Action
Figure
1
Rotate the exhausted or clogged sterile filter in a
clockwise position to remove it.
2
Unpack the new sterile filter and screw it into the
bottom 1/4“ thread on the sampling tap.
1/4“ thread
0.2 µm final filter
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...