1-4
7000740r7 Cryo Series
Thermo Scientific
Section 1
Quick Start-Up
Included with each unit (Model 746/747) is a lid strap for user
convenience.
1. Remove the protective white nylon screws from the areas indicated.
Discard these screws.
2. Install the strap as shown below, using the screws included with the
strap.
Warning
Use only a grounded electrical receptacle. Failure to ground the
unit can result in serious injury from electrical shock.
▲
With the power switch turned OFF, connect the unit to a grounded
electrical outlet. See the tag on the back of the unit or the electrical
schematics included in this manual for voltage and full load amps.
The power switch on the back of the unit is the
mains disconnect
and is
also a reset-type circuit breaker. If an overload condition occurs, the built-
in circuit breaker will trip and the power switch will turn off. Turning the
power switch on resets the circuit breaker. If the circuit breaker trips again
within a short time period, the unit should be checked by a qualified
electrician.
Connect Lid Strap
Figure 1-3.
Lid Strap
Connect Electrical
Supply
Summary of Contents for Cryo 740 Series
Page 17: ......
Page 23: ......
Page 27: ......
Page 31: ......
Page 39: ......
Page 40: ...7000740r7 Cryo Series 7 1 Thermo Scientific Section 7 Electrical Schematics ...
Page 41: ...7 2 7000740r7 Cryo Series Thermo Scientific Section 7 Electrical Schematics ...
Page 42: ...7000740r7 Cryo Series 7 3 Thermo Scientific Section 7 Electrical Schematics ...
Page 43: ...7 4 7000740r7 Cryo Series Thermo Scientific Section 7 Electrical Schematics ...
Page 50: ...7000740r7 Cryo Series 10 1 Thermo Scientific ...
Page 51: ......