Removing the Rotor
Thermo Scientific
75003667 BIOLiner
3-3
4. Close the centrifuge door.
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge door.
2. Grab the rotor handle with both hands and press against the green AutoLock
™
button. At the
same time, pull the rotor directly upwards with both hands and remove it from the centrifuge
spindle. Make sure not to jam the rotor while doing this.
Figure 3-2.
Handling of AutoLock
™
Aerosol-tight Rotors
CAUTION
Check that the rotor is properly locked on the centrifuge spindle before
each use by pulling its handle.
Be sure to check all seals before starting any aerosol-tight applications.
CAUTION
Rotors supplied with a lid for aerosol-tight applications come with a
mandril, which belongs to the AutoLock
™
. Be sure not to place the lid onto this
mandril to prevent it from being damaged.