1
Rotor Specifications
Technical Data
1-10
75003667 BIOLiner
Thermo Scientific
Table 1-9.
100 V, 50 / 60 Hz, Rectangular buckets 75003668
Centrifuge
Sorvall Legend XT
Sorvall Legend XF
Catalog #
75004508
75004535
Weight empty [kg]
7.3
7.3
Max. cycle number
55000
55000
Maximum permissible load [g]
4 x 750
4 x 750
Maximum speed n
max
[rpm]
3500
3500
Maximum RCF value at n
max
2739
2739
Max. / min. radius [cm]
20 / 10.5
20 / 10.5
Pitch angle [°]
90
90
Accel. / braking time [s]
75 / 60
75 / 60
Sample heating at n
max
[° C] referred to
ambient temperature of 23-25 °C, running
time 120 minutes
8
8
Aerosol-tight
*
yes
yes
Permissible temperature range autoclavable
(cycle number) °C
121
121
*
Tested by HPA, Porton-down, UK
Centrifuge
Sorvall Legend XTR
Sorvall Legend XFR
Catalog #
75004523
75004541
Weight empty [kg]
7.3
7.3
Max. cycle number
55000
55000
Maximum permissible load [g]
4 x 750
4 x 750
Maximum speed n
max
[rpm]
3500
3500
Maximum RCF value at n
max
2739
2739
Max. / min. radius [cm]
20 / 10.5
20 / 10.5
Pitch angle [°]
90
90
Accel. / braking time [s]
70 / 60
70 / 60
Minimum temperature at n
max
[°C] relating
to ambient temperature 23-25 °C
2
2
Speed at 4 °C
3500
3500
Aerosol-tight
*
yes
yes
Permissible temperature range autoclavable
(cycle number) °C
121
121
*
Tested by HPA, Porton-down, UK