Setting the Cell Constant
The actual temperature and cell constants are attached to
the cell cable internal to the Nanopure unit. This data
helps ensure that the purity displayed is as accurate as
possible. The only time the user would need to enter this
data is if the cell or main board is being replaced and/or
updated.
1.
Shut power OFF to the unit and disconnect the
power cord.
2.
Unlatch and open the right side door.
3.
Find and copy down the two cell constants
attached to the gray cable, leaving the electron-
ics board mounted on the inside of the right side
door.
4.
Close and latch the right side door.
5.
Reconnect the power cord and turn unit ON.
6.
From the (Idle) mode, when the display reads,
“Nanopure (Idle) xx:xx:xx xx/xx/xx,” press the
UP or DOWN arrow until the display reads,
“Advanced Menu.”
7.
Press ENTER.
8.
Proceed through the options by pressing
ENTER until “Set Cell Constant” is displayed.
9.
Press the UP or DOWN arrow to select “Yes.”
10. Press ENTER.
11.
Note that the digits are entered from left to right.
12. Press the UP or DOWN arrow to change the
“blinking” digit. Once the digit is correct, press
ENTER to advance to the next digit.
13. After all digits have been entered, line 2 of the
display will read, “(NO) Finished.”
14. If the cell constants displayed on line 1 of the
display are correct, press the UP or DOWN
arrow to select “Yes.”
U
SER
S
ETTINGS
25