AQ3170 Chlorine Colorimeter PROFORMA User Guide

Menu Selections

Switch the unit off.
Press the [MODE] key and hold .
Switch the unit on using the [ON/OFF] key . 

  

Allow the 3 decimal points to be displayed before releasing the [MODE] key .
The [Option] key allows for selection of the following menu points:
diS 

recall stored data

dAtE  setting the date and time
CAL  user calibration

diS  –  Recall of Stored Data

After confirming the selection with the [MODE] key the photometer shows the last 16 data sets 
in the following format (automatically proceeds every 3 seconds until result is displayed):
Number  

n xx (xx: 16 . . .1)   

Year 

 

YYYY (e .g . 2014) 

 

Date   

mm .dd (monthmonth:dayday)    

Time   

hh:mm (hourhour:minuteminute)   

Test  

 

Method   

Result  

x,xx

The [ZERO/TEST] key repeats the current data set .

The [MODE] key scrolls through all stored data sets .

Quit the menu by pressing [Option] key .

Setting Date and Time (24-hour-format)

After confirming the selection with the [MODE] key the value to be edited will be shown for 2 sec.

The setting starts with the year (YYYY) followed by the actual value to be edited . The same 
applies for month (mm), day (dd), hour (hh) and minutes (mm). Set the minutes first in steps of 
10, press the [Option] key to continue setting the minutes in steps of 1 .

Increase the value by pressing the [MODE] key .

Decrease the value by pressing [ZERO/TEST] key .

Proceed to the next value to be edited by pressing [Option] key .
After setting the minutes and pressing the [Option] key the display will show “IS SET” and the 
instrument returns to the measurement mode .

On
Off

Mode

Option

Mode

Mode

 Zero

 Test

diS

CAL

METHOD

0.0.0

METHOD

METHOD

CAL

CAL  –  User Calibration

Note:

user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)

After confirming the selection with the [MODE] key the instrument will show CAL/"Method".

Scroll through methods using the [MODE] key .

Fill a clean vial with the standard up to the 10 ml mark, screw the cap on and place the vial in 
the sample chamber making sure that the    marks are aligned .
Press the [ZERO/TEST] key .

The method symbol flashes for approx. 8 seconds.

The display shows the following in alternating mode:

Perform calibration with a standard of known concentration (see "Operation") .

Press the [ZERO/TEST] key .

The method symbol flashes for approx. 3 seconds.

The result is shown in the display, alternating with CAL .

If the reading corresponds with the value of the calibration standard (within the specified toler-
ance), exit calibration mode by pressing the [ON/OFF] key .

Changing the displayed value:

Pressing the [MODE] key once increases the displayed value by 1 digit .

Pressing the [ZERO/TEST] key once decreases the displayed value by 1 digit .

Press the corresponding key until the reading equals the value of the calibration standard .

By pressing the [ON/OFF] key, the new correction factor is calculated and stored in the user 
calibration software .

Confirmation of calibration (3 seconds).

CAL

cAL

 Zero

 Test

 Zero

 Test

Mode

SET

DATE

YYYY

 

(2 sec.)

Mode

 Zero

 Test

CAL

x

CAL

Mode

On
Off

 Zero

 Test

:             :

RESULT

dAtE

Cal

CAL

Option

Option

Methods

Menu options  -  Calibration Mode

Summary of Contents for AQUAfast AQ3170

Page 1: ...AQUAfast AQ3170 Chlorine Colorimeter User Guide Version 1 XCALI 97414 Revision A February 2017 ...

Page 2: ... worn out batteries and accumu lators They must not be disposed of in normal domestic waste Because our products include batteries and accumulators in the delivery package our advice is as follows Used batteries and accumulators are not items of domestic waste They must be disposed of in a proper manner Your local authority may have a disposal facility alternatively you can hand them in at any sho...

Page 3: ...lt in wrong readings Check at regular intervals and if necessary clean the transparent cell chamber using a moist cloth or cotton buds 9 Large temperature differences between the instrument and the environment can lead to errors e g due to the formation of condensation in the cell chamber or on the vial 10 To avoid errors caused by stray light do not use the instrument in bright sunlight 11 The re...

Page 4: ... the vial and place in the sample chamber making sure that the marks are aligned Press the ZERO TEST key For Countdown reaction period see page 7 The CL L or CL H symbol flashes for approx 3 seconds The result appears in the display The result is saved automatically Repeating the test Press the ZERO TEST key again Repeating the zero Press the ZERO TEST key for 2 seconds On Off Mode Zero Test Zero ...

Page 5: ...mbined Chlorine Total Chlorine Free Chlorine Tolerances 0 1 mg l 0 05 mg l 1 2 mg l 0 10 mg l Zero Test CL L CL L CL L Notes 1 Vial cleaning As many household cleaners e g dishwasher detergent contain reducing substances the subsequent determination of chlorine may show lower results To avoid any measurement errors only use glassware free of chlorine demand Preparation Put all applicable glassware...

Page 6: ...ement errors only use glassware free of chlorine demand Preparation Put all applicable glassware into sodium hypochlorite solution 0 1 g l for one hour then rinse all glassware thoroughly with deionised water 2 For individual testing of free and total chlorine the use of different sets of glassware is recommended EN ISO 7393 2 5 3 3 Do not use the same sample vial for free and total chlorine witho...

Page 7: ...tion key the display will show IS SET and the instrument returns to the measurement mode On Off Mode Option Mode Mode Zero Test diS CAL METHOD 0 0 0 METHOD METHOD CAL CAL User Calibration Note user calibration Display in calibration mode factory calibration Display in calibration mode After confirming the selection with the MODE key the instrument will show CAL Method Scroll through methods using ...

Page 8: ...ro Test RESULT Cal Technical Data Instrument single wavelength direct reading colorimeter Light source LED interference filter IF and photosensor in transparent cell chamber Wavelength specifications of the IF 530 nm 5 nm Wavelength accuracy 1 nm Photometric accuracy 3 FS T 20 C 25 C Photometric resolution 0 01 A Power supply 4 batteries AAA LR 03 Operating time 17hr operating time or 5000 test me...

Page 9: ...alists Email wai techservbev thermofisher com Within the United States call 1 800 225 1480 Outside the United States call 1 978 232 6000 or fax 1 978 232 6031 For additional product information contact your local authorized dealer Thermo Scientific Orion technical sales represen tative or contact us using the Water and Laboratory Products WLP information on the back page of this user manual Orderi...

Page 10: ......

Page 11: ...ll Free 1 800 225 1480 Tel 1 978 232 6000 info water thermofisher com Germany Tel 49 6184 90 6000 info water uk thermofisher com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asia thermofisher com Australia Tel 613 9757 4300 in Australia 1300 735 295 InfoWater...

Reviews: