AQ3170 Chlorine Colorimeter PROFORMA User Guide

General Notes

Guidelines for photometric measurements

1 .  Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent interference . 

Even minor reagent residues can cause errors in the test result .

2 .  The outside of the vial must be clean and dry before starting the analysis . Clean the outside of the vials 

with a towel to remove fingerprints or other marks.

3 .  Zero calibration and test must be carried out with the same vial as there may be slight differences in 

optical performance between vials .

4 .  The vials must be positioned in the sample chamber for zeroing and test with the   mark  on the vial 

aligned with the   mark on the instrument .

5 .  Always perform zeroing and test with the vial cap tightly closed . Only use the cap with a sealing ring .
6 .  Bubbles on the inside wall of the vial lead to incorrect measurements . To prevent this, remove the bubbles 

by swirling the vial before performing the test .

7 .  Avoid spillage of water into the sample chamber because this can lead to incorrect test results .
8 .  Contamination of the transparent cell chamber can result in wrong readings . Check at regular intervals 

and – if necessary – clean the transparent cell chamber using a moist cloth or cotton buds .

9 .  Large temperature differences between the instrument and the environment can lead to errors – e.g. due 

to the formation of condensation in the cell chamber or on the vial .

10 . To avoid errors caused by stray light do not use the instrument in bright sunlight .
11 . The reagents must be added in the correct sequence .

Method Notes

•  Prior to measurement ensure that the sample is suitable for analysis (no major interferences) and does 

not require any preparation i.e. pH adjustment, filtration etc.

•  Reagents are designed for use in chemical analysis only and should be kept well out of the reach of children .
•  Ensure proper disposal of reagent solutions .
•  Safety Data Sheets are available on request .

Correct filling of the vial:

Correct position of the vial (Ø 24 mm):

CAUTION:

To ensure that the instrument is water proof:
seal ring (E) must be in position
battery compartment cover (B) must be fixed with the four screws
If the batteries are removed for more than one minute the date and time menu starts automatically when the photo-
meter is switched on the next time .

Battery Replacement:

AA

A/L

R0

3

AA

A/L

R0

3

AA

A/L

R03

AA

A/L

R03

(D) batteries

(E) seal ring

(C) notch

(F) instrument back

(B) battery com-

partment cover

(A) screws

General Notes

Summary of Contents for AQUAfast AQ3170

Page 1: ...AQUAfast AQ3170 Chlorine Colorimeter User Guide Version 1 XCALI 97414 Revision A February 2017 ...

Page 2: ... worn out batteries and accumu lators They must not be disposed of in normal domestic waste Because our products include batteries and accumulators in the delivery package our advice is as follows Used batteries and accumulators are not items of domestic waste They must be disposed of in a proper manner Your local authority may have a disposal facility alternatively you can hand them in at any sho...

Page 3: ...lt in wrong readings Check at regular intervals and if necessary clean the transparent cell chamber using a moist cloth or cotton buds 9 Large temperature differences between the instrument and the environment can lead to errors e g due to the formation of condensation in the cell chamber or on the vial 10 To avoid errors caused by stray light do not use the instrument in bright sunlight 11 The re...

Page 4: ... the vial and place in the sample chamber making sure that the marks are aligned Press the ZERO TEST key For Countdown reaction period see page 7 The CL L or CL H symbol flashes for approx 3 seconds The result appears in the display The result is saved automatically Repeating the test Press the ZERO TEST key again Repeating the zero Press the ZERO TEST key for 2 seconds On Off Mode Zero Test Zero ...

Page 5: ...mbined Chlorine Total Chlorine Free Chlorine Tolerances 0 1 mg l 0 05 mg l 1 2 mg l 0 10 mg l Zero Test CL L CL L CL L Notes 1 Vial cleaning As many household cleaners e g dishwasher detergent contain reducing substances the subsequent determination of chlorine may show lower results To avoid any measurement errors only use glassware free of chlorine demand Preparation Put all applicable glassware...

Page 6: ...ement errors only use glassware free of chlorine demand Preparation Put all applicable glassware into sodium hypochlorite solution 0 1 g l for one hour then rinse all glassware thoroughly with deionised water 2 For individual testing of free and total chlorine the use of different sets of glassware is recommended EN ISO 7393 2 5 3 3 Do not use the same sample vial for free and total chlorine witho...

Page 7: ...tion key the display will show IS SET and the instrument returns to the measurement mode On Off Mode Option Mode Mode Zero Test diS CAL METHOD 0 0 0 METHOD METHOD CAL CAL User Calibration Note user calibration Display in calibration mode factory calibration Display in calibration mode After confirming the selection with the MODE key the instrument will show CAL Method Scroll through methods using ...

Page 8: ...ro Test RESULT Cal Technical Data Instrument single wavelength direct reading colorimeter Light source LED interference filter IF and photosensor in transparent cell chamber Wavelength specifications of the IF 530 nm 5 nm Wavelength accuracy 1 nm Photometric accuracy 3 FS T 20 C 25 C Photometric resolution 0 01 A Power supply 4 batteries AAA LR 03 Operating time 17hr operating time or 5000 test me...

Page 9: ...alists Email wai techservbev thermofisher com Within the United States call 1 800 225 1480 Outside the United States call 1 978 232 6000 or fax 1 978 232 6031 For additional product information contact your local authorized dealer Thermo Scientific Orion technical sales represen tative or contact us using the Water and Laboratory Products WLP information on the back page of this user manual Orderi...

Page 10: ......

Page 11: ...ll Free 1 800 225 1480 Tel 1 978 232 6000 info water thermofisher com Germany Tel 49 6184 90 6000 info water uk thermofisher com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asia thermofisher com Australia Tel 613 9757 4300 in Australia 1300 735 295 InfoWater...

Reviews: