Table 11
Storage information for the SeqStudio
™
Genetic Analyzer Cartridge v2 (Cat. No. A41331)
Condition
Description
Shipping
Shipped at 2–8°C.
Store upright at 2–8°C upon receipt.
Save the white storage box and optical cover for off-instrument cartridge storage.
On-instrument storage
For routine use, can be used and stored on the instrument for up to 6 months. If
you store the cartridge on-instrument:
• The instrument must be powered on.
• A Cathode Buffer Container must also be installed.
The instrument keeps the components under the following conditions when it is
powered on and in Cartridge storage mode:
• Optical detection window—Covered
• Capillary array electrodes—Submerged in cathode buffer (buffer must be
above the fill line in the Cathode Buffer Container)
• Polymer—Chilled
• Anode buffer—Ambient temperature
IMPORTANT!
The instrument does not maintain the correct temperature
conditions for the cartridge when it is powered off. Avoid cartridge exposure to
ambient temperature.
Off-instrument storage
For intermittent use, can be stored off-instrument until the expiry date on the label
or up to 6 months after first use. Store upright at 2–8°C, with an integrated
capillary protector (ICP) and optical cover installed (see “Store the cartridge“ on
Note:
After you remove the cartridge from the instrument, install an ICP within a
few minutes. Avoid cartridge exposure to ambient temperature.
Reuse
Can be removed from an instrument then inserted again on the same instrument
or a different instrument, if it was stored properly at 2–8°C and has not expired or
exceeded 250 injections.
Information about the cartridge installation and usage is retained in the cartridge
history (Settings
4
Cartridge
4
Instrument–cartridge history).
If a cartridge is installed on the instrument, the instrument will enter Cartridge
storage mode after the time you specify.
The instrument keeps the components under the following conditions when it is
powered on and in Cartridge storage mode:
• Optical detection window—Covered
• Capillary array electrodes—Submerged in cathode buffer (buffer must be above
the fill line in the Cathode Buffer Container)
• Polymer—Chilled
• Anode buffer—Ambient temperature
In the home screen:
Set cartridge
storage mode
(administrator
only)
Chapter 11
Maintain the instrument
Manage the cartridge
11
SeqStudio
™
Genetic Analyzer Instrument and Software User Guide
159