3.
Grasp the cartridge above the
capillaries.
4.
Position the cartridge:
a.
Orient the cartridge with the
embossed arrow pointing
toward the rear of the
instrument.
b.
Align the guides at the top of
the cartridge with the insertion
rails in the instrument.
5.
Slide the cartridge into the
instrument until it clicks into place.
6.
Close the instrument door, then
touch Done when the Consumables
Status screen is displayed.
The cartridge is initialized.
1.
Touch , touch Eject cartridge, then open the instrument door when
prompted.
2.
Hold the cartridge at the hand holds above the capillaries, then pull to remove it
from the instrument.
3.
Close the instrument door.
4.
Carefully place the optical cover on
the cartridge to avoid damage to the
capillaries.
2
1
1
Insertion rails
2
Embossed arrow
Store the
cartridge
1
1
Optical cover
Chapter 11
Maintain the instrument
Manage the cartridge
11