Accela 600 Pump or Accela 1250 Pump Installation
Solvent Line Connections
Thermo Scientific
Accela 600 and 1250 Pump Hardware Manual
29
Figure 15.
Flat and angled sides of the steel compression ring
To connect the solvent lines to the degasser inlets of the Accela pump
1. Insert the color-coded Super Flangeless fitting into the appropriate inlet port
(see
), and then tighten the nut fingertight. Be careful not to cross-thread the
fitting, which can cause solvent leakage.
• For inlet A, use the blue fitting.
• For inlet B, use the yellow fitting.
• For inlet C, use the green fitting.
• For inlet D, use the orange fitting.
Figure 16.
Vacuum degasser connections
2. To complete the installation, route the solvent line through the cutouts along the left side
of each Accela module.
Tip
The Super Flangeless fittings are color coded to match the color-coded solvent
composition graphic in the pump program area of the data system.
Angled side
Flat side
Vacuum degasser
inlets and outlets