3
AutoLock
TM
Removing the Rotor
Thermo Scientific
75003603 BIOShield
TM
1000A
3-3
4. Close the centrifuge door.
Removing the Rotor
To remove the rotor, proceed as follows:
1. Open the centrifuge door.
2. Grab the rotor handle with both hands and press against the green AutoLock
TM
button. At the
same time, pull the rotor directly upwards with both hands and remove it from the centrifuge
spindle. Make sure not to jam the rotor while doing this.
CAUTION
Check that the rotor is properly locked on the centrifuge spindle before each use
by pulling it a its handle.
Be sure to check all seals before starting any aerosol-tight applications.