8-4
12A Water Still
Thermo Scientific
Section 8
Troubleshooting
Demineralizer Troubleshooting
Problem
Causes
Solution
Still will not operate.
Cartridge expended. Red light
burned out. Relay contacts
open. No power
Change cartridge. Check light
& replace. If operational, clean
contacts. If not, replace. Check
fuse.
No green light.
Green light burned out.
Check & replace.
Still on with red light on
demineralizer.
Relay contacts closed. Input
jack not plugged in.
Check and replace relay.
Check and reconnect jack.
Both lights on,or good lights
will not light, or erratic light
operation.
Printed circuit board.
Replace.
Automatic Collection System Troubleshooting
Problem
Causes
Solution
Will not turn still on or off as
it should.
Float stuck.
Remove level control cover
and free-up float.
Switch not working. Input jack
not plugged in.
Replace switch. Check and
reconnect jack.