Use the following procedure when shutting down the instrument.
1.
Connect a clean air source to the SAMPLE connection and allow the
instrument to purge pollutants for at least 10 minutes.
2.
Turn the power OFF.
3.
Cover instrument gas ports when not in use.
IMPORTANT TIPS
●
Do not pressurize the analyzer with sample.
●
Do not allow liquids to enter the sample port.
●
Do not expose the instrument to vibration or magnetic interference.
●
Do not remove/adjust the small circular entrance or exit mirrors in the
optical bench.
●
Use an external pump with atmospheric bypass dump to deliver the
sample gas to the instrument.
●
Filter/condition sample stream. Sample dew point must be between
2
°
C and 8
°
C.
●
Operate the instrument in a climate controlled environment
(15-30
°
C).
●
Set DHCP to OFF when changing the IP address. If DHCP is ON,
the instrument will respond with “Not Settable if DHCP is ON.” For
more information on DHCP, see the “Operation” chapter in the
“Model 60
i
Instruction Manual.”