Thermo Scientific 60i Quick Start Manual Download Page 5

Thermo Fisher Scientific 

Model 60

i

 Quick Start Guide     

5

  

 

 

 

 

Figure 4.

 Rear Panel Connectors 

 

Data acquisition and external control are implemented through Ethernet, 
RS-232 or a dedicated I/O system and analog outputs. If a dedicated I/O 
system will be used, connections are made through terminal block PCB 
assemblies with a D-Sub connector on one side and a series of screw 
terminals on the other. Refer to the “Installation” chapter in the “Model 
60

i

 Instruction Manual” for detailed information on I/O configuration. 

Refer to Figure 5 and Table 2 for information on analog outputs and the 
most commonly used digital inputs. Refer to Table 3 for information on 
optional analog current outputs and optional analog voltage outputs (I/O 
expansion board 25-pin connector.)   

The following terminal block PCB assemblies are available for iSeries 
instruments:  

 

I/O terminal block PCB assembly, 37 pin (standard) 

 

D/O terminal block PCB assembly, 37 pin (standard) 

 

25-pin terminal block PCB assembly (included with optional I/O 
Expansion Board) 

 

 

 

 

Making Data and 

I/O Connections

Component Side Viewed from Top of Board 

24 

12 

13 

Terminal Block

AC Power 

Fan

61i Calibrator Connector 
(optional) 

Standard I/O 

Optional 4-20mA Current Outputs 
and Analog Voltage Inputs 

Plumbing Connectors

62i Probe Control 
(optional) 

Ethernet 

RS-232/485 

Digital Outputs

Summary of Contents for 60i

Page 1: ...Model 60i Quick Start Guide Multi Gas Analyzer Part Number 106474 00 23Jul2009 Example of a Direct Extractive Continuous Emissions Monitoring System CEMS ...

Page 2: ...ep Action 1 Connect the sample line from the chiller bypass pump to the SAMPLE bulkhead on the rear panel Ensure that the sample line is not contaminated by dirty wet or incompatible materials All tubing should be constructed of FEP Teflon with an OD of 1 4 inch Note Sample gas must be filtered and dried when delivered to the analyzer The filter should have a nominal pore size of 1 micron or less ...

Page 3: ...pan gases and zero air to CAL IN via inch tubing 4 Connect the EXHAUST bulkhead to a suitable vent The exhaust line should be inch OD with an ID of 1 8 inch or greater The length of the exhaust line should be less than 10 feet Verify that there is no restriction in this line Note If calibration gases will not be introduced through the sample probe see the Calibration chapter in the Model 60i Instr...

Page 4: ...be Barrel Calibration gas connections to manifold All calibration gases are regulated to the same pressure Usually 10 to 30 psi depending on choice of capillary or needle valve used downstream Calibration Gas Manifold Capillaries or needle valve restrict flow to 1 5 to 2 0 LPM for direct calibration or 5 to 6 LPM for system calibration Heated Sample Line Calibration Line Heated Particulate Filter ...

Page 5: ... analog outputs and the most commonly used digital inputs Refer to Table 3 for information on optional analog current outputs and optional analog voltage outputs I O expansion board 25 pin connector The following terminal block PCB assemblies are available for iSeries instruments I O terminal block PCB assembly 37 pin standard D O terminal block PCB assembly 37 pin standard 25 pin terminal block P...

Page 6: ...7 2 4 6 8 10 12 Signal Grounds NA 1 13 Power Fail NC NA 20 14 Power Fail Com NA 21 15 Power Fail NO NA 3 16 TTL Input 1 Turn Pump Off 4 17 TTL Input 2 Trigger Blow Back Pin 1 Pin 20 Pin 3 Digital Input Channel 1 Pin 4 Digital Input Channel 2 Pin 23 Digital Input Channel 3 Pin 24 Digital Input Channel 4 Pin 6 Digital Input Channel 5 Pin 25 Digital Input Channel 6 Pin 7 Digital Input Channel 7 Pin 2...

Page 7: ...e All channels are user definable If any customization has been made to the analog output configuration or TTL inputs the default selections may not apply Defaults shown are for a fully equipped analyzer Table 3 Optional Analog Current Outputs and Analog Voltage Inputs I O Expansion Board 25 Pin Connector Rear Panel Connector Pin Terminal Block Screw Number Signal Description 15 1 IOut1 13 2 Isola...

Page 8: ...The Model 60i is supplied with a three wire grounded power cord Under no circumstances should this grounding system be defeated Use the following procedure when starting the instrument 1 Turn the power ON and allow two hours for the instrument to stabilize 2 Set instrument parameters such as operating ranges and averaging times to appropriate settings For more information about instrument paramete...

Page 9: ...not expose the instrument to vibration or magnetic interference Do not remove adjust the small circular entrance or exit mirrors in the optical bench Use an external pump with atmospheric bypass dump to deliver the sample gas to the instrument Filter condition sample stream Sample dew point must be between 2 C and 8 C Operate the instrument in a climate controlled environment 15 30 C Set DHCP to O...

Page 10: ...rights reserved Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin MA 02038 1 508 520 0430 www thermo com aqi ...

Page 11: ... describes possible alarm messages that may be displayed during start up and provides the recommended action to take to restore the system to normal operation Troubleshooting Alarm Messages Alarm Message Possible Cause Action Sample Flow Pump defective Pump disconnected Defective or disconnected flow switch cable Improper alarm threshold setting Plumbing connection configuration Repair or replace ...

Page 12: ... Intensity Analog Input Readings Analog Input Voltages Digital Inputs Relay States Test Analog Outputs Instrument Config Contact Info Alarms Concentration Alarms Other Alarms System Alarms Service Absorbances Show User Zero Show User Span Pressure Calibration O Sensor Calibration Preamp Board Cal Temperature Calibration Analog Out Cal Lock Analog Outputs Analog Out Lock Type Set Analog Lock Percen...

Reviews: