background image

Replacing Parts

Speed Control:

Replacement of the speed control requires removal of the

bottom cover casting. Turn unit over and remove 6

screws holding cover and lift off.

To replace speed control, remove the control knob (2

setscrews) and a retaining nut to separate it from the

panel. Unplug the terminal connections from the old ter-

minal unit and plug them into the new one. Reverse this

sequence to put the control unit back into position.

The new control unit must be calibrated in the following

manner. Locate the maximum and minimum speed

potentiometers on the speed control. Lay the unit on its

side and elevate the housing so that the titer plate plat-

form will clear the table to permit an unobstructed area of

shaking motion.

Start the unit and turn the speed control knob fully clock-

wise (maximum speed). Using a suitable tachometer and

a small plastic screwdriver, adjust the maximum speed

potentiometer to 1100 rpm.

Next, rotate the speed control fully counterclockwise (min-

imum speed). Now adjust the "minimum" potentiometer to

the point at which platform stops moving.

If the procedure as outlined has been followed carefully,

the new speed control is now calibrated.

Replace bottom cover casting and insert and securely

tighten the 6 screws that were previously removed.

Silicone Tubing Sections:

To replace the silicone tubing sections on the platform,

remove the 4 screws holding the platform and lift off.

Silicone rubber tubing is available from most laboratory

supply houses or from Thermo Scientific. Use 1/4" ID x

7/16" OD and cut to 1-5/16" length.

Reassemble the unit carefully by following the reverse

order from the preceding.

11

Maintenance

Note

Make no attempt to service or repair

a Thermo Scientific product under

warranty before consulting your dis-

tributor. After the warranty period,

such consultation is still advised,

especially when the repair may be

technically sophisticated or difficult.

Warning

Disconnect plug from electrical out-

let before attempting any

maintenance or repair of the unit.

To avoid risk of electric shock, dis-

connect equipment from power

source before removing the bottom

cover or performing any service.

Warning

Do not operate shaker without plat-

form. To do so will produce a serious

imbalance and cause the entire unit

to walk. In addition, exposed rotating

parts will be a hazard to fingers.

Summary of Contents for 4625

Page 1: ...9 3 10 ...

Page 2: ...Carton 8 Leveling 8 Electrical Requirements 8 Operation 9 Controls 9 Loading the Platform 9 Starting Stopping the Shaker 10 Setting the Speed Control 10 Power Mains Protection 10 Maintenance 10 Replacing Parts 11 Speed Control 11 Silicone Tubing Sections 11 Motor Brushes 12 Replacement Parts 13 Ordering Procedures 14 Warranty 15 ...

Page 3: ...l prior to the use of this equipment Warnings To avoid electrical shock always 1 Use a properly grounded electrical outlet of cor rect voltage and current handling capacity 2 Disconnect from the power supply prior to main tenance and servicing To avoid personal injury 1 Do not use in the presence of flammable or combustible materials or explosive gases Do not use in the presence of pressurized or ...

Page 4: ...ds or make the unit inoperative 9 Perform regular maintenance service as speci fied in this manual and keep unit in good repair Do not operate with known defects 10 Make sure all vessels are securely clamped before turning on unit To eliminate hazard of electrical shock make sure floor around machine is dry In the event of accidental spilling or splashing of liquids clean up and or neutral ize the...

Page 5: ...OTOR Continuous duty rated high efficiency DC motor features high torque full ball bearing construction and ceramic permanent magnets Brushes are user replaceable SHAKING SPEED Speed range is 40 to 1100 rpm TIMER Range Up to 5 minutes in 1 minute increments or continuous run operation SHAKER MOTION 0 120 inch orbit DIMENSIONS 11 W x 11 D x 7 H 28 x 28 x 18 cm SHIPPING WEIGHT 25 lbs 11 kg ...

Page 6: ...Thermo Fisher Scientific Robert Bosch Strasse 1 D 63505 Langenselbold Germany ...

Page 7: ... speeds from 40 to 1100 rpm to produce shaking move ment ranging from gentle rotation to vigorous agitation to meet specific user requirements For increased versatility a rocker switch on the control panel enables users to select either a constant shaking motion or a timed cycle up to 5 minutes for accurate repetitive runs The ribbed platform provides a non skid surface and retaining springs which...

Page 8: ... damage within a few days after receipt of shipment absolves the carrier from any liability for damage You must call for a damage inspection promptly Leveling Place the shaker on a level surface in a convenient loca tion and near a properly grounded electrical outlet that matches the requirements of the unit as indicated on the nameplate A minimum of 1 2 inch clearance is required on all sides of ...

Page 9: ...sert a filled titer plate in the spring tensioned holder by pushing the plate against the springs until the titer plate is resting on the ribbed bottom of the platform and is securely locked in position 9 Operation Warning Do not use in the presence of flammable or combustible materi als or explosive gases Do not use in the presence of pressurized or sealed containers Fire or explo sion may result...

Page 10: ...he center or OFF position Setting the Speed Control After starting the shaker turn the speed control clockwise to the desired level of shaking action It is not necessary to return the speed control to the start or zero position after completion of each shaking cycle Once a satisfactory shak ing speed is found the control may be left at that setting To stop the shaker turn the switch to OFF If the ...

Page 11: ...s moving If the procedure as outlined has been followed carefully the new speed control is now calibrated Replace bottom cover casting and insert and securely tighten the 6 screws that were previously removed Silicone Tubing Sections To replace the silicone tubing sections on the platform remove the 4 screws holding the platform and lift off Silicone rubber tubing is available from most laboratory...

Page 12: ...bottom casting to expose the motor 2 brushes are located under plastic caps on opposite sides of the lower part of the motor Unscrew the plastic caps and slide the brushes out Replace brushes when they are worn down to 3 16 in length Always replace both brushes even if only one brush is worn down 12 MAINTENANCE ...

Page 13: ...72 00 Fuse 5x20 4 A 4625 1CE 2 FZX53 Motor Brush Set 370 277 00 Platform Mat 790 316 01 Retaining Ring 11 16 500 339 00 Retaining Spring 2 850 083 00 Rubber Feet 4 790 078 00 Speed Control 4625 227 841 00 Speed Control 4625 1CE 229 419 00 Speed Control Knob 560 223 00 Switch 440 373 00 Timer 4625 270 122 00 Timer 4625 1CE 270 129 00 Tube Restraining 3 790 298 01 Wave Washer 564 354 00 Wiring Schem...

Page 14: ...800 943 2006 or 800 926 0505 ...

Page 15: ...0 Italy 39 02 02 95059 434 254 375 Netherlands 31 76 571 4440 Nordic Baltic countries 358 9 329 100 Russia CIS 7 812 703 42 15 Spain Portugal 34 93 223 09 18 Switzerland 41 44 454 12 12 UK Ireland 44 870 609 9203 Asia China 86 21 6865 4588 or 86 10 8419 3588 India toll free 1800 22 8374 India 91 22 6716 2200 Japan 81 45 453 9220 Other Asian countries 852 2885 4613 Countries not listed 49 6184 90 6...

Reviews: