Functions
You can enter programming mode by pressing and holding for 5 seconds.
Pressing repeatedly scrolls through the available setpoint functions: cold
alarm and warm alarm.
Table 3.
Setpoint Programming Functions
Function
Programming Sequence
Adjust temperature
control setpoint
Enter programming mode by pressing
and holding for 5 seconds. On release, the
current temperature setpoint value flashes in the temperature display; use and to
adjust it. Press to confirm the new value. The display automatically returns to normal
operating mode 30 seconds after the last key entry or after scrolling through all available
functions.
Adjust cold alarm
setpoint
Enter programming mode by pressing
and holding for 5 seconds. On release, the
current cold alarm setpoint value flashes in the temperature display; use and to
adjust it. The display automatically returns to normal operating mode 30 seconds after
the last key entry or after scrolling through both available functions.
Adjust warm alarm
setpoint
Enter programming mode and press
repeatedly until the top of the thermometer is
illuminated. The current warm alarm setpoint value then flashes in the temperature
display; use and to adjust it. The display automatically returns to normal operating
mode 30 seconds after the last key entry or after scrolling through both available
functions.