8-2
High Performance Incubator
Thermo Scientific
Section 8
Servicing Guide
Replacing Temp
Controller
1. Place ON/OFF switch in OFF position.
2. Unplug incubator from outlet power supply.
3. To remove controller from control housing:
a) Use both hands to firmly grip each side of the controller bezel.
b) Press on the bezel side grips until the bezel tabs release.
c) Slowly pull controller from housing.
Note
PCB contacts at rear of controller fit into contacts at rear of
controller housing.
s
4. To install new, factory-configured controller:
a) Carefully slide new controller into controller housing.
b) Press controller bezel into controller housing until bezel tabs
securely lock controller into place.
5. Plug incubator into outlet power supply.
6. Place ON/OFF switch in ON position.