18  |   Maintenance and Care

Centrifuge GT 2/2R

Disinfection

Disinfect the centrifuge immediately whenever infectious 
material has spilled during centrifugation.

The rotor chamber and the rotor should be treated preferably 
with a neutral disinfectant.

Contact the Service Department of Fisher Scientific for 
questions regarding the use of other disinfectants.

Disinfect the rotor and accessories as follows:

1. Open the centrifuge.

2. Turn off the centrifuge.

3. Pull out the power supply plug.

4. Grasp the rotor with both hands and lift it vertically off the 

centrifuge spindle.

5. Remove the centrifuge tubes and adaptors and dispose 

of them or disinfect them.

6. Treat the rotor and accessories according to the 

instructions for the disinfectant (soak in solution). Adhere 
strictly to the given application times.

7. Be sure that disinfectant can drain off the rotor.

8. Rinse the rotor and accessories thoroughly with water.

9. Dispose of the disinfectant according to the applicable 

guidelines.

10. Dry all of the rotors and accessories after cleaning with a 

cloth or in a warm air cabinet at a maximum temperature 
of 50 °C.

After cleaning, treat the entire surface of aluminum parts 
with corrosion protection oil (70009824). Also treat the 
cavities with oil.

Tread the bold of the swing out rotor with bold grease 
(75003786).

Decontamination

Decontaminate the centrifuge immediately whenever 
radioactive material has spilled during centrifugation.

For general radioactive decontamination use a solution of 
equal parts of 70% ethanol, 10% SDS and water. 

1. Open the centrifuge.

2. Turn off the centrifuge.

3. Pull out the power supply plug.

4. Grasp the rotor with both hands and lift it vertically off the 

centrifuge spindle.

5. Remove the centrifuge tubes and adaptors and dispose 

of them or disinfect them.

6. Rinse the rotor first with ethanol and then with de-ionized 

water.

7. Adhere strictly to the given application times.

8. Be sure the decontamination solution can drain off the 

rotor.

9. Rinse the rotor and accessories thoroughly with water.

10. Dispose of the decontamination solution according to the 

applicable guidelines.

11. Dry all of the rotors and accessories after cleaning with a 

cloth or in a warm air cabinet at a maximum temperature 
of 50 °C.

After cleaning, treat the entire surface of aluminum parts 
with corrosion protection oil (70009824). Also treat the 
cavities with oil.

WARNING: 

Infectious material can get into the 

centrifuge when a tube breaks or as a result of 
spills. Keep in mind the risk of infection when 
touching the rotor and take all necessary 
precautions.

In case of contamination, make sure that others 
are not put at risk.

Decontaminate the affected parts immediately.

Take other precautions if need be.

CAUTION: 

Before using any cleaning or 

decontamination methods except those 
recommended by the manufacturer, users 
should check with the manufacturer that the 
proposed method will not damage the 
equipment.

Observe the safety precautions and handling 
instructions for the cleaning agents used.

WARNING: 

Radioactive material can get into 

the centrifuge when a tube breaks or as a result 
of spills. Keep in mind the risk of infection when 
touching the rotor and take all necessary 
precautions.

In case of contamination, make sure that others 
are not put at risk.

Decontaminate the affected parts immediately.

Take other precautions if need be.

CAUTION: 

Before using any cleaning or 

decontamination methods except those 
recommended by the manufacturer, users 
should check with the manufacturer that the 
proposed method will not damage the 
equipment.

Summary of Contents for 15808722

Page 1: ...Fisherbrand GT2 Centrifuge Fisherbrand GT2R Centrifuge Instruction manual 50156459_a 04 2018 15808722 15818722 ...

Page 2: ...onal purposes only The contents and the product it describes are subject to change without notice Fisher Scientific makes no representations or warranties with respect to this manual In no event shall be held liable for any damages direct or incidental arising from or related to the use of this manual BIOShield Fiberlite HIGHConic Microliter MicroClick and Auto Lock All trademarks are the property...

Page 3: ...nical Data 3 Directives Standards and Guidelines 4 Functions and Features 4 Mains Supply 5 Refrigerant 5 Rotor Selection 6 Before Use 7 Before Setting up 7 Transporting the Centrifuge 7 Location 7 Aligning the Centrifuge 7 Mains Connection 8 Storage 8 Control Panel 9 Control Panel 9 Keys 9 Operation 10 Switch on Centrifuge 10 Lid Opening 10 ...

Page 4: ... Rotor 13 Aerosol tight Rotors 14 Aligning the Centrifuge 14 System Menu 15 Description 16 Maintenance and Care 17 Cleaning Intervals 17 Cleaning 17 Disinfection 18 Decontamination 18 Autoclaving 19 Service of Fisher Scientific 19 Troubleshooting 20 Mechanical Emergency Door Release 20 Troubleshooting by User 21 When to Contact Customer Service 24 Chemical Compatibility Chart 25 ...

Page 5: ...ntrifuge should be operated by trained specialists only The centrifuge is to be used for its intended use only Plug the centrifuge only into sockets which have been properly grounded Do not move the centrifuge while it is running Do not lean on the centrifuge Use only rotors and accessories for this centrifuge which have been approved by Thermo Scientific Exceptions to this rule are commercially a...

Page 6: ...fuging microbiological samples from the Risk Group II according to the Bio safety Manual of the World Health Organization WHO aerosol tight biological seals have to be used For materials in a higher risk group extra safety measures have to be taken If toxins or pathogenic substances have gotten into the centrifuge or its parts appropriate disinfection measures have to be taken Refer to section Dis...

Page 7: ... is equipped with a view port and a lock The lid of the centrifuge can only be opened while the centrifuge is switched on and the rotor has come to a complete stop The centrifuge cannot be started until the lid has been closed properly The integrated rotor detection systems ensures that no inadmissible speed settings can be preselected Electronic imbalance recognition Lid emergency release For eme...

Page 8: ...10 mm Weight without rotor 91 5 kg 57 5 kg Table 1 Technical Data Centrifuge GT 2 2R Feature Values Table 2 Directives Standards and Guidelines Tension Frequency Directives Produced and inspected according to the following standards and guidelines 220 230 V 50 60 Hz 2006 42 EC Machinery 2014 35 EU Low Voltage Protective Goals 2014 30 EC Electromagnetic Compatibility EMC 2011 65 EC RoHS Directive o...

Page 9: ... RCF selection temperature control pretemp with cooling equipped devices Acceleration braking profiles 9 acceleration and 10 braking curves Rotor recognition Automatic Imbalance recognition Electronic contingent on rotor and speed Lid lock Automatic lid closing and locking starting from an initial hold position Table 3 Functions and Features Component Function Description Features Table 4 Electric...

Page 10: ...e found in the corresponding rotor operating manuals For more information visit our website at http www thermofisher com TX 400 75003629 with round buckets 75003655 TX 200 75003658 with round buckets 75003659 BIOShield 720 75003621 M 20 75003624 Microliter 30x2 sealed 75003652 CLINIConic 30x15 75003623 8x50 sealed 75003694 FIBERLite F15 6x100 75003698 HIGHConic II 75003620 TX 1000 TX 750 BIOShield...

Page 11: ...ed around the centrifuge People and hazardous substances must be kept out of the safety zone while centrifuging The supporting structure must be stable and free of resonance for example a level laboratory bench The supporting structure must be suitable for horizontal setup of the centrifuge The centrifuge should not be exposed to heat and strong sunlight The set up location must be well ventilated...

Page 12: ...s of your country 4 Make sure that the voltage and frequency correspond to the figures on the rating plate 5 Establish the connection to the power supply with the connecting cable Storage Before storing the centrifuge and the accessories it must be cleaned and if necessary disinfected and decontaminated Store the centrifuge in a clean dust free location Be sure to place the centrifuge on its feet ...

Page 13: ...ULSE key Temperature Speed RCF value Acceleration curve Braking curve Bucket selection OPEN key Key Display Contents Start Normal start of the centrifuge Stop End run manually Open lid Automatic release possible only when device is switched on Emergency release see Mechanical Emergency Door Release Pulse By pressing the PULSE key the centrifuge starts immediately and accelerates up to the end spee...

Page 14: ... only the rotors and accessories from this list in the centrifuge The centrifuge is equipped with an Auto Lock locking system This system is used to automatically lock the rotor to the centrifuge spindle The rotor does not have to be bolted on to the centrifuge spindle Proceed as follows 1 Open the lid of the centrifuge and if necessary remove any dust foreign objects or residue from the chamber A...

Page 15: ... the key repeatedly until the desired value shows Explanation of RCF Value The relative centrifugal force is given as a multiple of the force of gravity g It is a unit less numerical value which is used to compare the separation or sedimentation capacity of various devices since it is independent of the type of device Only the centrifuging radius and the speed come into play in it r centrifuging r...

Page 16: ...alue shows 3 The display shows The display shows the current temperature inside the rotor chamber 4 Press the START key The rotor chamber is cooled down or heated up to the preset temperature 5 Press the STOP key The display shows the current temperature inside the rotor chamber Saving programs 1 Enter the program parameters 2 Press any of the program store keys for 4 seconds Centrifugation Once t...

Page 17: ...lid will open and you can remove the centrifuged material Temperature Adaptation during Standstill The temperature cannot be adapted until the rotor has been positively identified the speed display will then show END When the rotor is not recognized lid closed and START key not yet pressed speed display 0 the centrifuge responds by ensuring that the sample cannot freeze regardless of the rotor bei...

Page 18: ...uto Lock lid for aerosol tight rotors Aligning the Centrifuge To turn off the centrifuge put the mains switch to 0 CAUTION Rotors supplied with a lid for aerosol tight applications come with a mandrel which belongs to the Auto Lock Be sure not to place the lid onto this mandrel to prevent it from being damaged WARNING Mind the Auto Lock mandrel inside the lid Do not touch the mandrel ...

Page 19: ...tart Stop Start Language Enter system menu End of run beep no Keypad beep no LCD powersave no Auto lid open no Software ID 0547V01 Cycle count 2368 Press the STOP key to exit the system menu English English Deutsch Francaise Español Italiano Nederlands Росский 0547V01 4641V051 1 Software and NVRAM identification example values yes no ...

Page 20: ...trifuge should make beep when pressing any key Otherwise use the keys below the bucket selection until it says NO Use the keys below the speed display in order to navigate through the system menu Press the STOP key to quit the system menu LCD Powersave Use the keys below the bucket selection until it says YES in the display if the centrifuge should enter a powersave mode after the run Otherwise us...

Page 21: ... centrifuge spindle 5 Remove the centrifuge tubes and adapters 6 Use a neutral cleaning agent with a pH value between 6 and 8 for cleaning 7 Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50 C After cleaning treat the entire surface of aluminum parts with corrosion protection oil 70009824 Also treat the cavities with oil Tread...

Page 22: ...off the centrifuge spindle 5 Remove the centrifuge tubes and adaptors and dispose of them or disinfect them 6 Rinse the rotor first with ethanol and then with de ionized water 7 Adhere strictly to the given application times 8 Be sure the decontamination solution can drain off the rotor 9 Rinse the rotor and accessories thoroughly with water 10 Dispose of the decontamination solution according to ...

Page 23: ...hnician The service technicians check the following The electrical equipment The suitability of the set up site The lid lock and the safety system The rotor The fixation of the rotor and the drive shaft Fisher Scientific offers inspection and service contracts for this work Any necessary repairs are performed for free during the warranty period and afterwards for a charge This is only valid if the...

Page 24: ...ollows 1 Make sure the rotor has stopped view port in the lid 2 Pull out the power supply plug 3 On the left side of the housing is one white plastic plugs which you can pry out of the side plate with a screwdriver or a knife Pull the release cord attached to it to trigger the mechan ical lid release The lid will open and the samples can be removed Figure 5 Emergency Door Release 4 Push the cord b...

Page 25: ... centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked Rotor code is not in the rotor table It is permitted to use the rotor currently mounted in this device Restart the centrifuge If an error message appears again inform Customer Service Rotor ID failure The centrifuge cannot be operated The run does not start or the centrifuge runs down without bei...

Page 26: ...entrifuge runs down without being braked Error during the self test of the centrifuge program and the electronics Restart the centrifuge If an error message appears again inform Customer Service E 13 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked The check sum in the data memory is incorrect The software corrects errors automatically Check...

Page 27: ... operated The run does not start Check whether you selected the right bucket Is it easy to turn the rotor when the lid is open Does the rotor rub against the device Restart the centrifuge If an error message appears again inform Customer Service E 30 The centrifuge cannot be operated The run does not start or the centrifuge runs down without being braked Restart the centrifuge If an error message ...

Page 28: ...fuge cannot be operated The run does not start or the centrifuge runs down without being braked The centrifuge accelerates too slowly Check whether you selected the right bucket Check whether you selected the right bucket Is it easy to turn the rotor when the lid is open Does the rotor rub against the device Restart the centrifuge If an error message appears again inform Customer Service E 41 E 74...

Page 29: ...Alcohol U S S S S M S S S M S S Aluminum Chloride U U S S S S U S S S S M S S S S S S S S S M U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate M S U S S S S S S S S S S S U U S S S S S S M S S S Ammonium Hydroxide 10 U U S U S S M S S S S S S U M S S S S S S S S S M S Ammoniumhydroxid 28 U U S U S U M S S S ...

Page 30: ...l Ether S S U U S S S U S U U S U U U U U U U U U S S S S M U Diethyl Ketone S U U M S U S M U U U M M U S S U U Diethylpyrocarbonate S S U S S S S S U S U S U S S S M S S S S S S Dimethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U Dioxane M S U U S S M M S U U S U M U U M M M U S S S S U U Ferric Chloride U U S M S M S S S S M U S S Acetic Acid Glacial S S U U S S U M S U S U U U ...

Page 31: ... S S S S M M M S Iodoacetic Acid S S M S S S S M S S M S S M S S S S S M S S M M Potassium Bromide U S S S S S S S S S S S S S S S S S S S M S S S Potassium Carbonate M U S S S S S S S S S S S U S S S S S S S S S S S S Potassium Chloride U S S S S S S S S S S S S S S S S S S S S U S S S Potassium Hydroxide 5 U U S S S S M S S S S S U S S S S S S S M U M S U Potassium Hydroxide conc U U M U M M S S...

Page 32: ... U S M M S U S N N Dimethylformamide S S S U S M S S S U S U S U U S S U U S M S S S U Sodium Borate M S S S S S S S S S S U S S S S S S S S S S M S S S Sodium Bromide U S S S S S S S S S S S S S S S S S S S M S S S Sodium Carbonate 2 M U S S S S S S S S S S S S U S S S S S S S S S S S S Sodium Dodecyl Sulfate S S S S S S S S S S S S S S S S S S S S S S S S Sodium Hypochlorite 5 U U M S S M U S S ...

Page 33: ... S S S S S S S S S S S M S S S Rubidium Chloride M S S S S S S S S S S S S S S S S S S M S S S Sucrose M S S S S S S S S S S S S S S S S S S S S S S S S S Sucrose Alkaline M S S S S S S S S S S S U S S S S S S S S M S S S Sulfosalicylic Acid U U S S S S S S S S U S S S S S S S S S U S S S Nitric Acid 10 U S U S S U U S U S U S S S S S S S S S M S S S S Nitric Acid 50 U S U M S U U S U S U U M M U ...

Page 34: ... S S S S M S U S Hydrogen Peroxide 3 S M S S S S S S S S S S S S M S S S S S S S S S S Xylene S S U S S S M U S U U U U U U M U M U U U S U M S U S Zinc Chloride U U S S S S U S S S S S S S S S S S S S S S S U S S S Zinc Sulfate U S S S S S S S S S S S S S S S S S S S S S S S S S Citric Acid 10 M S S M S S M S S S S S S S S S M S S S S S S S S S S CHEMICAL MATERIAL ALUMINUM ANODIC COATING for ALUM...

Page 35: ... smaltito o riciclato tramite queste medesime Ulteriori informazioni sulla conformità di Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodotti Scientific che possono essere utili alla rilevazione di sostanze soggette alla Direttiva RoHS sono disponibili sul sito www thermofisher com WEEERoHS in Servizi e Supporto Conformité WEEE Ce ...

Page 36: ...870 241 1034 unity ie thermofisher com Italy 39 02 950 59 478 it fishersci com 39 800 823 162 unity it thermofisher com Finland 358 0 9 8027 6280 fishersci fi 46 8 556 468 20 unity fi thermofisher com France 33 0 3 88 67 14 14 fr fishersci com 33 1 60 92 49 50 unity fr thermofisher com Netherlands 31 0 20 487 70 00 nl fishersci com 31 0 76 579 55 55 unity bnl thermofisher com Norway 47 22 95 59 59...

Page 37: ...Thermo Electron LED GmbH Branch Osterode Am Kalkberg 37520 Osterode am Harz Germany Find out more at eu fishersci com 50156459_a 15808722 15818722 ...

Reviews: