background image

Dionex PDA Operator’s Manual

14

Doc. 065378-01 1/12

2.3

Optical System

Figure 2-4

 is a schematic of the Dionex PDA optical system. 

Light from the tungsten lamp is focused through an opening in the internal 
structure of the deuterium lamp. Light from the tungsten and deuterium lamps is 
then focused through the flow cell by the source lens. After exiting the cell, the 
light passes through the spectrograph lens to the filter paddle and is focused into 
the slit. The light then passes through the slit to the grating, where it is diffracted 
into the component wavelengths. Measurement of the light occurs at the 
photodiode array. Each diode measures a narrow portion of the spectrum. 

Figure 2-4.  Dionex PDA Optical System

Grating

Slit

Flow Cell

Filter Paddle

Photodiode Array

Spectrograph Lens

Source Lens

Visible Focus Lens

Deuterium Lamp

Tungsten Lamp

Summary of Contents for 074114

Page 1: ...Dionex ICS Series Photodiode Array Detector Operator s Manual Document No 065378 Revision 01 January 2012...

Page 2: ...to use in the product operation This document is copyright protected and any reproduction of the whole or any part of this document is strictly prohibited except with the written authorization of The...

Page 3: ...verview 6 1 4 Safety and Regulatory Information 7 1 4 1 Safety Messages and Notes 7 1 4 2 Safety Labels 9 1 4 3 Declaration of Conformity 9 2 Description 11 2 1 Front Features 11 2 2 Inside Front Pane...

Page 4: ...Connect to the Chromeleon 7 Client 34 3 1 3 Connect to the Chromeleon 6 8 Client 35 3 1 4 Turn On the Lamps 37 3 2 System Equilibration 37 3 3 Routine Operation 38 3 3 1 Direct Control 38 3 3 2 Automa...

Page 5: ...rge or Small 84 4 13 Poor Peak Shape 84 4 14 Faulty USB Communication 86 4 14 1 Moduleware Run Time Diagnostics 87 4 14 2 Software Diagnostics and Calibration 87 5 Service 91 5 1 Liquid Leaks 91 5 2 C...

Page 6: ...rs 108 B Installation 111 B 1 Facilities Requirements 111 B 2 Unpacking 112 B 3 Installing the Flow Cell 113 B 4 Connecting the Waste Line 116 B 5 Connecting the Analog Outputs Optional 116 B 6 Connec...

Page 7: ...the USB Device Drivers 124 B 10 Configuring the Dionex PDA in Chromeleon 125 B 10 1 Using the USB Auto Configuration Wizard 125 B 10 2 Selecting Configuration Properties 127 B 11 Starting the Chromel...

Page 8: ...Dionex PDA Operator s Manual vi Doc 065378 01 1 12...

Page 9: ...Dionex PDA can be used in numerous laboratory environments for routine analysis as well as for sophisticated research tasks A deuterium lamp optimizes the UV range 190 to 380 nm and a tungsten lamp o...

Page 10: ...eon 6 8 Chromatography Data System Chromeleon 7 and Chromeleon 6 8 also provide data acquisition and data processing functions NOTE Thermo Scientific Dionex Chromeleon Xpress software can provide real...

Page 11: ...be programmed and all compounds that absorb within this range can be detected in a single analysis Determination of the correct wavelengths in one analysis After all peaks have been detected the maxi...

Page 12: ...ity obtained from two consecutive chromatographic analyses Based on 2D data or if the analyses had been performed with a conventional variable wavelength detector the peaks might have been misidentifi...

Page 13: ...1 Introduction Doc 065378 01 1 12 5 Figure 1 3 Spectra of Pyrene and an Impurity Impurity Pyrene...

Page 14: ...duction to the chromatography software Chromeleon required for detector control Chapter 3 Operation and Maintenance Instructions for routine Dionex PDA operation with Chromeleon and guidelines for opt...

Page 15: ...Canada call the nearest Thermo Fisher Scientific office 1 4 1 Safety Messages and Notes This manual contains warnings and precautionary statements that can prevent personal injury and or damage to the...

Page 16: ...ation de danger potentiel qui si elle n est pas vit e pourrait entra ner des blessures mineures mod r es galement utilis pour signaler une situation ou une pratique qui pourrait gravement endommager l...

Page 17: ...PDA is in compliance with the following standards EN 61010 1 2001 safety CAN CSA C22 2 No 1010 1 92 A2 97 safety UL 61010C 1 2002 R8 02 safety and EN 61326 1997 A1 1998 EMC susceptibility and immunity...

Page 18: ...Dionex PDA Operator s Manual 10 Doc 065378 01 1 12...

Page 19: ...tions see Figure 2 2 Use the power button for routine on off control of the Dionex PDA NOTE The main power switch is on the Dionex PDA rear panel see Figure 2 8 Always leave the main power switch on u...

Page 20: ...amp is firing The LED flashes every 0 5 second if a command is issued to turn on the lamp but the lamp fails to turn on see Section 4 3 If a subsequent attempt to turn on the lamp is successful the LE...

Page 21: ...flow cell is located behind a cover leaks from the cell tubing connections will not be apparent immediately To prevent damage to detector components the bottom of the optical bench contains a built i...

Page 22: ...the source lens After exiting the cell the light passes through the spectrograph lens to the filter paddle and is focused into the slit The light then passes through the slit to the grating where it i...

Page 23: ...tainless steel semi micro cells in both PEEK and 316 stainless steel and semi preparative cells in PEEK only Spectrograph Lens Receives the light from the flow cell and focuses it onto the slit Filter...

Page 24: ...PL Up to 1 0 mL min SEMI PREPARATIVE FLOW CELL Cell Material Cell Volume Cell Path Length Volume of Heat Exchanger Inlet Tubing Intended Flow Rate Range PEEK P N 064167 0 7 PL 0 4 mm N A 5 0 to 100 0...

Page 25: ...with normal phase or chlorinated solvents these solvents will damage the cell Do not use a stainless steel flow cell with low pH solutions these solutions may cause corrosion contamination and metal l...

Page 26: ...l is designed for applications with 2 to 3 mm ID separator columns and flow rates up to 1 0 mL min Figure 2 6 Semi Micro Flow Cell Assembly Semi Micro Cell Component PEEK Cell SST Cell Inlet tubing 0...

Page 27: ...h semi preparative flow applications and flow rates of 5 0 to 100 0 mL min The 0 4 mm path length minimizes detector absorbance for concentrated peaks Figure 2 7 Semi Preparative Flow Cell Assembly Ce...

Page 28: ...l utilisateur Toute r paration doit tre effectu e par un personnel qualifi utilisant des proc dures correctes de d charge lectrostatique Halten Sie sich von der Elektronik des Dionex PDA fern Die Ele...

Page 29: ...mp board Preamp output is directed to the Data Processing board Data Processing Board The Dionex PDA Moduleware resides on the Data Processing board This board provides the digital signal processing a...

Page 30: ...analog output settings will be either the factory defaults see Section 3 3 3 or the settings selected the last time the detector was controlled by the computer Drain Tube Connector Waste exits the de...

Page 31: ...N 060977 includes one USB cable P N 063246 For installation instructions see Section B 7 4 Power Entry The Dionex PDA power entry is socketed for a modular power cord IEC 320 C13 The detector operates...

Page 32: ...illez ce que la prise de base soit situ e install e pr s du module et facilement accessible STROMSCHLAGGEFAHR Zur Vermeidung von elektrischen Schl gen ist eine geerdete Steckdose zu verwenden Das Ger...

Page 33: ...advantageous to locate the Dionex PDA ahead of the suppressor This location also limits suppressor exposure to backpressure When changing from a buffer to a different operating mobile phase be sure th...

Page 34: ...olvents methanol moderate concentrations of acetonitrile etc are acceptable for a PEEK system Other solvents and high concentrations of acetonitrile require a stainless steel system All tubing connect...

Page 35: ...matography Data System Both software products provide extensive instrument control data management reporting and compliance features In Chromeleon 7 an ePanel Set provides centralized system control Y...

Page 36: ...378 01 1 12 functions The UV_Vis tab provides access to Dionex PDA functions as well as detailed status and diagnostics information Figure 2 9 shows the Dionex PDA ePanel in Chromeleon 7 Figure 2 9 Ex...

Page 37: ...rol commands are executed as soon as they are entered See Section 3 3 for details about direct control With automated control you create a Chromeleon 7 instrument method or a Chromeleon 6 8 program An...

Page 38: ...on license can be reviewed and processed on any Chromeleon client station 2 7 3 System Wellness and Predictive Performance System Wellness monitors the overall health of the chromatographic system It...

Page 39: ...c 065378 01 1 12 31 Leak detection Wavelength verification System Wellness and Predictive Performance commands and parameters are available in the Chromeleon 7 Command window or the Chromeleon 6 8 Com...

Page 40: ...Dionex PDA Operator s Manual 32 Doc 065378 01 1 12...

Page 41: ...h a Dionex ICS 5000 or Dionex ICS 3000 system use the power button on the front of the Dionex Dual Pump DP or Dionex Single Pump SP for routine on off control To turn off the pump press and hold the p...

Page 42: ...he taskbar click Start All Programs Chromeleon 7 Services Manager to open the Services Manager and click Start Instrument Controller 2 Start the Chromeleon 7 client a Click Start on the Windows taskba...

Page 43: ...rver right click the icon and select Start Server When the server is running and data acquisition is not occurring the icon is gray If the Server Monitor icon is not on the taskbar click Start on the...

Page 44: ...Control panel select the Dionex PDA tab on the panel tabset see Figure 3 1 d Verify that the Connected check box is selected If it is not click the box to connect the Dionex PDA to Chromeleon 6 8 NOTE...

Page 45: ...o the temperature required for the application Set the detector wavelength s and turn on the lamp s Calibrate the wavelength s Monitor the pump pressure and verify that the reading is correct for the...

Page 46: ...ions refer to the Chromeleon Help 3 3 1 Direct Control With direct control you select operating commands and parameters from the following locations Chromeleon 7 ePanel Sets or the Command window or C...

Page 47: ...n of Chromeleon installed The display filter level Normal Advanced or Expert The options selected for the detector in the Properties dialog box in the Chromeleon 7 Instrument Configuration program or...

Page 48: ...lect the desired filter on the context menu To open the Chromeleon 6 8 Commands dialog box 1 Open the Dionex PDA Control panel 2 Select Control Commands or press the F8 key The Commands dialog box app...

Page 49: ...the main steps required to set up and run the detector automatically Refer to the Chromeleon Help for details about these steps NOTE Before creating a program review Section 3 4 for a description of...

Page 50: ...settings are at the factory defaults listed in the table below These default settings are also in effect when the power is first turned on Rise time 2 0 seconds UV lamp On Visible lamp On Stand Alone...

Page 51: ...ivity vs baseline noise Reference wavelength Baseline drift Reference bandwidth Baseline noise baseline drift Bunch width Spectral resolution peak match disk space Step Baseline noise rise time data c...

Page 52: ...e signal and results in less short term noise see Figure 3 5 However a rise time that is too long may result in reduced peak heights and asymmetric peak shapes When set correctly the rise time signifi...

Page 53: ...few seconds select a faster data collection rate for example 5 0 Hz If the data collection rate is too slow the start and end points of peaks are not accurately determined If the collection rate is to...

Page 54: ...ace it provides complete information about the analytes and the spectra can be used to determine peak identity purity and wavelength maxima Bandwidth The bandwidth defines a wavelength range around th...

Page 55: ...nex PDA The selected wavelength will be used as the reference wavelength The reference intensity will vary over time as the signal intensity changes For maximum compensation select a reference wavelen...

Page 56: ...is ideal for gradient analyses where the baseline may shift This mode also helps compensate for fluctuations in ambient temperature caused by air conditioning To cancel the active reference mode ente...

Page 57: ...dwidth 10 nm Reference bandwidth 50 nm active reference mode only Flowing water at 1 0 mL min 10 35 MPa 1500 psi Figure 3 7 Effect of Reference Wavelength on Baseline Drift Reference Wavelength Mode O...

Page 58: ...ut may increase the chance that the sample absorbs in the reference bandwidth this will reduce sensitivity The reference wavelength plus one half the reference bandwidth must be within the sample wave...

Page 59: ...he required data storage by approximately the reciprocal Do not select a bunch width when the spectrum of the compound has a fine structure the bunch width will reduce the spectral resolution Figure 3...

Page 60: ...selected for each data collection rate Step Override In almost all cases the step setting selected by Chromeleon 6 8 is the preferred setting However it is possible to override this setting by manual...

Page 61: ...en step intervals reports that value and noise is reduced This setting is recommended for most applications When Average is Off Chromeleon 6 8 reports the data points at the step interval The data poi...

Page 62: ...execute the command to calibrate the intensity of the deuterium lamp Afterward check the lamp intensity again If the reading is still too low clean the flow cell see Section 5 2 or replace the lamp s...

Page 63: ...e controlled devices While the system is in standby mode the detector lamps remain on Both processes maintain the equilibrated status of an instrument until the system is restarted 3 6 Routine Mainten...

Page 64: ...Steel Systems and Components Document No 031152 Monitor baseline noise If the noise increases above an acceptable level especially after a lamp s has been in service for an extended length of time re...

Page 65: ...wavelength verification test for both the deuterium and tungsten lamps approximately every 6 months see Section 4 14 2 If a lamp fails the test calibrate the wavelength see Section 4 14 2 and then run...

Page 66: ...Dionex PDA Operator s Manual 58 Doc 065378 01 1 12...

Page 67: ...anada call the nearest Thermo Fisher Scientific office 4 1 Alarms and Error Conditions The Moduleware the instrument control firmware installed in the Dionex PDA periodically checks the status of cert...

Page 68: ...te calibration command during data acquisition Error page 65 Can t execute diagnostic command during data acquisition Error page 65 Can t execute x command during data acquisition Error page 65 Comman...

Page 69: ...ror while attempting to turn off Abort page 70 UV lamp error while attempting to turn on Abort page 70 UV lamp failed to turn on for the required operation Abort page 70 Visible lamp control circuitry...

Page 70: ...Wavelength 190 to 800 nm Reference wavelength 750 nm Reference bandwidth 50 nm These parameters are valid because 750 nm 25 nm 50 of 50 nm 775 nm which is less than 800 nm The following parameters are...

Page 71: ...s than 500 nm The following parameters are NOT VALID Wavelength 190 to 600 nm Reference wavelength 750 nm Reference bandwidth 50 nm These parameters are invalid because 190 to 600 nm is a range of mor...

Page 72: ...o troubleshoot This message appears if you attempt to select a parameter for the 3D field while data acquisition is in progress Wait until data acquisition is complete or stop the run and then change...

Page 73: ...when the computer attempts to communicate with the detector Shut down the Chromeleon instrument controller or server and turn off the Dionex PDA power wait 30 seconds and then power up both the instr...

Page 74: ...t Thermo Fisher Scientific for assistance A backup paddle alignment routine ensures that the Dionex PDA remains functional even if the sensor cannot detect the filter wheel position Thus if this is th...

Page 75: ...mo Fisher Scientific personnel To troubleshoot This message appears if you enter a rise time setting that is invalid for a data collection rate of 20 Hz Select a valid rise time To troubleshoot This m...

Page 76: ...id and dry the leak sensor before resuming operation If the cell leaks after you resume operation replace it see Section 5 6 To troubleshoot Follow the troubleshooting steps in Section 4 7 To troubles...

Page 77: ...r an invalid setting for the bunch width Select a bunch width setting from 1 to 25 nm To troubleshoot The Dionex PDA can measure the absorbance spectrum from 190 to 800 nm Make sure both the minimum a...

Page 78: ...ot pinched under the lamp cover 3 Check the number of hours the lamp has been in service Open the Chromeleon 7 Command window see page 31 or the Chromeleon 6 8 Commands dialog box see page 32 and then...

Page 79: ...f you notice a shift in the absorbance signal during this time replace the lamp If the error message relates to the tungsten lamp execute the VisLampAge command If the tungsten lamp has been in operat...

Page 80: ...st be performed by Thermo Fisher Scientific personnel To troubleshoot Reduce the data collection rate to less than 20 Hz or select a wavelength range that is d 310 nm UV lamp control circuitry error w...

Page 81: ...Does Not Light Lamp LED Is Flashing Rapidly Lamp default setting is not in effect When the factory default settings are in effect the deuterium and tungsten lamps turn on automatically each time the d...

Page 82: ...Setting or VisLampPowerOnSetting is On If the LED continues flashing replace the lamp See Section 5 7 or Section 5 8 for instructions on replacing the deuterium lamp or tungsten lamp respectively Dete...

Page 83: ...e applications work or for detector validation allow at least 2 hours for the lamps to stabilize After installing a new lamp allow from 8 to 24 hours for the lamp to stabilize After changing mobile ph...

Page 84: ...an 2000 hours If this occurs replace the lamp See Section 5 7 or Section 5 8 for instructions on replacing the deuterium lamp or tungsten lamp respectively Lamp incorrectly installed Make sure that bo...

Page 85: ...erest See page 44 for more information Detector exposed to high vibration Optical detectors are sensitive to vibrations Make sure the Dionex PDA installation site is vibration free Filter paddle has m...

Page 86: ...to temperature and light Do not operate the Dionex PDA unless the front cover lamp cover and flow cell cover are in place Flow cell heat exchanger not in place Check that the flow cell inlet tubing is...

Page 87: ...LED DEUTERIUM UV or TUNGSTEN VIS is lighted If necessary turn on one of the lamps from Chromeleon or the TTL input Low light throughput 1 Remove the cell from the optical bench see Section 5 4 2 Check...

Page 88: ...heck the Chromeleon Audit Trail If the Filter wheel sensor could not detect blocked position or Filter wheel sensor could not detect open position message is displayed see page 66 for instructions Fai...

Page 89: ...UV_Calibration c Execute the Intensity command to calibrate the lamp intensity d Execute the Autozero command to run the autozero routine e If no error message is displayed flush the cell with the mob...

Page 90: ...displayed flush the cell with deionized water or methanol until the autozero routine can be performed without error Lamp s needs replacement Check the number of hours the lamps have been in service 1...

Page 91: ...program or the Dionex PDA Control panel 4 11 Low Spectral Resolution Bunch width too wide Select a bunch width setting of 1 nm and then check the resolution again Bandwidth too wide Select a bandwidt...

Page 92: ...that the reagent flow rate is correct Reference bandwidth too wide Select a narrower reference bandwidth The reference bandwidth should not overlap the sample bandwidth 4 13 Poor Peak Shape Concentrat...

Page 93: ...elect a bandwidth that provides the resolution required to achieve the desired sensitivity A wider bandwidth reduces baseline noise but may increase the chance that the sample absorbs in the reference...

Page 94: ...solve the problem Click Cancel to exit the wizard Turn off the detector and unplug the USB cable from the computer Install Chromeleon Reconnect the USB cable to the computer and turn on the power to t...

Page 95: ...s and Calibration Chromeleon includes comprehensive diagnostics for the Dionex PDA This section describes some of the most frequently used diagnostic procedures For more information refer to the Chrom...

Page 96: ...cute the WavelengthCalibration command In Chromeleon 6 8 a Verify the following conditions no solvent is flowing through the cell the background absorbance is low and there are no bubbles in the light...

Page 97: ...bleshooting Doc 065378 01 1 12 89 In Chromeleon 6 8 a To open the Wellness panel click the Wellness button on the Dionex PDA Control panel b Click the wavelength verification script button under Diagn...

Page 98: ...Dionex PDA Operator s Manual 90 Doc 065378 01 1 12...

Page 99: ...roblem 5 1 Liquid Leaks The Dionex PDA is plumbed with the parts listed below For tightening requirements refer to Installation of Dionex Liquid Line Fittings Document No 031432 The manual is on the T...

Page 100: ...97 N A Cell outlet tubing 0 38 mm 0 015 in ID P N 057304 N A Cell inlet tubing 0 12 mm 0 005 in ID N A P N 063903 Cell outlet tubing 0 25 mm 0 010 in ID N A P N 051650 Reducing union fittings P N 0559...

Page 101: ...l with a succession of at least 10 PL each of deionized water acetone and 3 M HNO3 4 Reconnect the cell liquid lines 5 Flush the cell with deionized water for 20 to 30 minutes at 1 0 mL min 5 3 Removi...

Page 102: ...ides and pull straight off to remove 2 Disconnect the tubing connected to the flow cell inlet and outlet see Figure 5 1 3 Remove the flow cell cover from the inside front panel see Figure 2 3 4 Standa...

Page 103: ...low cell from the optical bench see Section 5 4 and place the cell on a clean work surface 2 Disassembling the cell with bare hands may introduce contaminants into the system Before proceeding put on...

Page 104: ...n turn an additional one quarter turn 9 Repeat Step 3 through Step 8 to install the second window 10 Connect the flow cell to the pump and pressurize to 1 38 MPa 200 psi Check for leaks around the win...

Page 105: ...ce stains the cell interior and cleaning the cell see Section 5 2 does not remove the stain or if excessive backpressure cracks the fused silica windows and you do not want to replace them see Section...

Page 106: ...amp emits UV radiation that is harmful to the eyes Always turn off the detector power before removing the lamp cover La lampe deuterium emettet de rayons ultraviolets qui sont dangeroux pour les yeux...

Page 107: ...dust If necessary clean the lamp with IPA isopropyl alcohol 12 Line up the flange of the new lamp with the two locating pins and three threaded holes in the optical bench see Figure 5 3 When the flan...

Page 108: ...troller or server 21 Reset the lamp lifetime a Open the Chromeleon 7 Command window see page 31 or the Chromeleon 6 8 Commands dialog box see page 32 b Select UV and then select the UVLampAge command...

Page 109: ...sible lamp connector and disconnect it from the 4 pin bulkhead connector on the side wall see Figure 5 3 10 Use a screwdriver to loosen the tungsten lamp screws and then pull the mounting assembly awa...

Page 110: ...reinstall the tubing bracket 16 Replace the flow cell cover and reconnect the inlet and outlet tubing 17 Replace the detector front cover 18 Turn on the detector power 19 Start the Chromeleon instrum...

Page 111: ...center to release it The fuse holder will pop out slightly when the lock is released Repeat for the other lock Pull the fuse holder straight out of the compartment HIGH VOLTAGE Disconnect the main po...

Page 112: ...r Scientific recommends always replacing both fuses 6 Reinsert the fuse holder into the compartment The fuse holder is keyed to fit only in the correct orientation Apply pressure evenly against the ho...

Page 113: ...r 100 W Maximum line draw 3 5 A at 110 VAC at power up Fuses Two fast blow IEC 127 fuses P N 954763 rated at 4 0 A Dimensions Height 17 4 cm 6 8 in Width 44 4 cm 17 5 in Depth 50 3 cm 19 8 in Clearanc...

Page 114: ...rative Cell Installed 10 PAU at 254 nm 2 second rise time 4 nm bandwidth and flowing water at 1 0 mL min 15 PAU at 520 nm 2 second rise time 10 nm bandwidth and flowing water at 1 0 mL min Noise with...

Page 115: ...use bases stronger than 0 1 M these solutions will etch the fused silica windows of the flow cell Do not use normal phase or chlorinated solvents with the PEEK flow cell these solutions will damage th...

Page 116: ...ns may cause corrosion contamination and metal leaching Cell Body PEEK Volume and Optical Path Length 0 7 PL volume 0 4 mm 0 02 in path length Maximum Operating Pressure 10 34 MPa 1500 psi Chemical Co...

Page 117: ...A Specifications Doc 065378 01 1 12 109 Volume of Semi Micro Cell Heat Exchangers including cell inlet tubing PEEK cell 5 PL Stainless steel cell 5 PL...

Page 118: ...Dionex PDA Operator s Manual 110 Doc 065378 01 1 12...

Page 119: ...bration free workbench of a height that allows convenient access to the interior of the Dionex PDA Allow at least 6 cm 2 4 in behind the Dionex PDA for power connections and ventilation Do not block a...

Page 120: ...place B 2 Unpacking Carefully remove the Dionex PDA from the shipping container Do not turn on the Dionex PDA unless it is at room temperature 4 to 40 C 40 to 104 F If you move the detector from a col...

Page 121: ...inlet tubing through the tubing guide on the lamp cover and reinstall the tubing bracket see Figure 5 1 6 Thread the flow cell outlet tubing through the slot on the side of the flow cell cover Replace...

Page 122: ...Standard or Semi Micro Flow Cell Connections Figure B 2 Semi Preparative Flow Cell Connections Cell Inlet Tubing Reducing Union Fitting PEEK cell P N 055903 Stainless steel cell P N 055902 Reducing U...

Page 123: ...ow rate of 1 0 mL min 2 meters for a flow rate of 0 5 mL min and so on c Install a 10 32 ferrule fitting and bolt on each end of the tubing d Using a union fitting P N 042627 connect the backpressure...

Page 124: ...ionex PDA rear panel see Figure B 3 The four Dionex PDA analog outputs provide a 1 volt output maximum The outputs are controlled by absorbance wavelengths 1 through 4 in Chromeleon The chromatography...

Page 125: ...put stage If the relay load can source more than 200 mA at 5V or higher it may damage the Data Processing board SHOCK AND FIRE HAZARD Inputs and outputs are not rated for connection to an AC power mai...

Page 126: ...ter The Chromeleon 7 or Chromeleon 6 8 license is installed NOTE Thermo Fisher Scientific strongly recommends installing Chromeleon before connecting the Dionex PDA to the computer When Chromeleon is...

Page 127: ...nt Controller The icon changes to gold to indicate that the Instrument Controller Service is starting When the Instrument Controller Service is running idle the icon changes to gray If the Chromeleon...

Page 128: ...directly to the PC Connect the detector to an external USB hub P N 060392 To connect the Dionex PDA directly to the PC 1 Plug the A connector of the USB cable P N 063246 into the USB port on the compu...

Page 129: ...1 Plug the A connector of the USB cable P N 063246 into the USB port on the module with the internal hub see Figure B 6 2 Plug the B connector of the USB cable into the USB receptacle on the Dionex PD...

Page 130: ...lug the B connector of the USB cable into a port on the external USB hub see Figure B 7 The USB standard limits the USB cable length to 5 meters 5 5 yds Each USB device can be separated from the PC by...

Page 131: ...t outlet is located near the Dionex PDA and is easily accessible Operation at AC input levels outside of the specified operating voltage range may damage the Dionex PDA DANGER D LECTROCUTION Pour vite...

Page 132: ...zard automatically appears Accept the default option for Chromeleon 7 Create a new instrument or Chromeleon 6 8 Create a new timebase and click OK This opens the Auto Configuration dialog box Leave th...

Page 133: ...d and then manually create an instrument or timebase For instructions on how to manually create an instrument refer to the Chromeleon 7 Help or Chromeleon 7 Installation Guide For instructions on how...

Page 134: ...ange Timebase Names dialog box appears now Enter a new and unique name and click the Use Custom Names button To cancel click the Use Default Names button NOTE Selecting an instrument or timebase name...

Page 135: ...roperties dialog box right click the detector in the instrument or timebase and click Properties The Properties dialog box for the detector appears see Figure B 8 2 Review the default settings for the...

Page 136: ...d displays the ePanel Set 3 To view the Dionex PDA ePanel see Figure B 9 click the UV_Vis tab on the ePanel Set The ePanel provides access to Dionex PDA functions as well as detailed status and diagno...

Page 137: ...ex PDA Control panel see Figure B 10 select the Dionex PDA tab on the panel tabset The Control panel provides access to Dionex PDA functions as well as detailed status and diagnostics information Figu...

Page 138: ...Dionex PDA Operator s Manual 130 Doc 065378 01 1 12...

Page 139: ...43276 Ferrule 043275 Nut 060498 Dionex AD25 Dionex PDA Dionex PDA 100 Flow Cell Windows Replacement Kit Standard Flow Cell 316 Stainless Steel 056126 Stainless steel standard flow cell with heat excha...

Page 140: ...exchangers 051650 Stainless steel semi micro flow cell outlet tubing assembly 055902 Reducing union fitting stainless steel 010262 Ferrule 010261 Nut 060498 Dionex AD25 Dionex PDA Dionex PDA 100 Flow...

Page 141: ...ture fluctuations 48 Analog output connectors 2 21 22 116 Default settings 42 Output 116 Analog output range 22 Analytes unknown 46 Audit Trail 13 59 Automated control 29 Average 43 Definition 53 Guid...

Page 142: ...n 7 2 3D Data Acquisition license 30 Audit Trail 13 59 87 Automated control 41 Command window 39 Diagnostic procedures 87 Direct control 38 Direct control mode 38 ePanel Set 27 Installation 119 Instru...

Page 143: ...anel Set 27 Equilibration system 75 Using the Smart Startup feature 38 Error messages 13 F Fan 21 68 Speed 21 Ferrule fittings 68 75 Part number 115 Filter paddle 14 15 Malfunction 77 80 Positions 15...

Page 144: ...nt of detector 12 Clearing the ALARM LED 73 Lenses 15 Liquid leaks 91 Liquid lines Maintenance 55 56 Lithium battery disposal 91 M Maintenance procedures 55 Mise en garde icon 8 Mobile phases 26 Conce...

Page 145: ...utdown procedure 55 Spare parts 131 Unpacking instructions 112 Warranty 91 Weight 105 Photodiode Array Panel ePanel 28 128 129 Post column reagents 84 Degassing 75 Power cord 23 103 123 POWER LED LED...

Page 146: ...cement procedure 97 See also Cells Volume 16 107 Windows replacement 95 96 Windows replacement kit 95 Semi preparative flow cells 15 Chemical compatibility 108 Flow rate range 16 Heat exchanger volume...

Page 147: ...83 Troubleshooting Baseline drift 78 79 Baseline noise 75 77 Elution time is slow 75 Lamp does not light 73 PDA does not respond 74 Peak shape is poor 84 Peak size is too big or small 84 Spectra not...

Page 148: ...8 W Warning icon 7 Warnung icon 8 Warranty voiding 91 Waste line 22 Installation 116 Wavelength calibration 2 82 88 Causes of failure 82 95 Wavelength verification 15 82 88 Wavelengths single 1 46 Ref...

Reviews: