95 Series Operator’s Guide
4-12
71-0011 — REV E
Troubleshooting Guide
The following troubleshooting guide lists possible indications,
probable cause, and recommended actions.
Table 4-1
95 Series Gas Monitor Troubleshooting Guide
Indication
Condition and
Probable Cause
Recommended Action
Blinking LOBAT
message. Buzzer
chirps and alarm
light flashes every
30 sec.
Low battery. Less than
8 hours of operating life
remaining.
Replace the battery.
Steady LOBAT
message. Steady
buzzer and alarm
light.
No gas reading.
Battery below operating
level. Instrument
inoperative.
Replace the battery immediately.
Steady FAULT
message. Steady
buzzer and alarm
light.
Bad sensor or faulty
sensor circuit.
Or, improper
calibration.
Recalibrate. If the indication
persists, replace the sensor and
recalibrate. If the indication
persists, contact the factory for
repair.
Blinking FAULT
message.
No buzzer or alarm
light.
Sensor is still
functioning but is near
the end of its life.
Recalibrate. If the indication
persists, replace the sensor as soon
as practical.
Calibrate the new sensor.
FAIL message
during calibration
procedure.
Improper calibration.
Or, sensor needs
replacement.
Repeat the entire calibration
procedure. If condition persists,
replace the sensor, then
recalibrate.