User Maintenance
______________________________________________________ API Source Maintenance
__________________ LCQ Deca XP Plus Hardware Manual _________________
4-27
Reinstalling the ESI Probe
To finish reinstalling the ESI probe into the ESI flange, install the ESI slide
back onto the ESI flange and bolt it down with four socket head cap screws.
Go on to the next topic: Reinstalling the ESI Probe Assembly.
Reinstalling the ESI Probe Assembly
Reinstall the ESI probe assembly on the XP Plus as described in the topic
Installing the ESI Probe Assembly in the Changing API Probe Assemblies
chapter, or place the ESI probe assembly in its storage container.
Maintaining the API Stack
The API stack includes the spray shield, ion sweep cone, spray cone, ion
transfer capillary, capillary lens and skimmer mount, and the tube lens, and
skimmer. The ion transfer capillary has a finite lifetime. You need to replace
the ion transfer capillary if the ion transfer capillary bore becomes corroded or
damaged. You also need to clean the spray shield, ion sweep cone, spray
cone, ion transfer capillary, tube lens, skimmer, and the other components of
the API stack on a periodic basis.
It is not necessary to vent the system in order to replace the ion transfer
capillary. To replace the ion transfer capillary, proceed as follows:
Figure 4-7.
Installing the ESI fused-silica sample tube
ESI
Manifold
Sample
Inlet
Fused-Silica Capillary
(Sample Tube)
Insert Sample Tube
From This Direction
ESI
Nozzle
ESI Probe
ESI Needle
1 mm
Sample Tube