11
c)
Press R/C for radiant chamber heat. Press at any time to turn OFF and
ON. (As long as there is time left in the heat timer).
d)
Once the time decrements to 0.00 HRS. the run will automatically stop,
the display will show “End”,the SAV valves will click, also allowing air to
bleed into the chamber.
e)
After the rotor stops spinning, the cover unlocks and the display reverts
to last set parameters, and the UNIT will sound several audible beeps.
f) Open the cover and remove samples.
GENERAL: During the run, display shows actual parameters. To check set parameters
press “VIEW” button and “SELECT”. The display will revert temporarily to set points.
3.5 OTHER TOPICS
Chamber cover.
The cover must be closed before beginning a run. If you
press “Manual Run” or “Auto Run” and the cover is open, the display shows
“lid” to remind you to close the cover.
A user cannot open the cover during a run due to vacuum in the
concentrator chamber. The cover is locked down at all times during a run
and whenever power to the unit is interrupted.
The cover lock is an additional safety feature that reduces the risk of injury
or damage from the spinning rotor. DO NOT bypass the cover lock to
conduct a run with the cover open. If removal of samples from the concen-
trator chamber during a power failure is required, insert a firm, thin object
into the hole on the side of the unit. Press the object gently into the hole
until the cover hook is released. The cover can then be opened.
In order to release the
cover, insert a firm,
thin object into this
hole and press the
object gently.