Thermo Electron Corporation
Polysonics SX40 User Guide
6-1
Chapter 6
Troubleshooting & Maintenance
The instrument has 4 modes of operation:
●
Setup
●
Learn
●
Operating
●
Manual (used by factory trained personnel primarily for application
troubleshooting)
Use the following table in conjunction with the learn mode and operating
mode warning tables.
Table 6–1.
Action codes
Action
Description
A
1. Turn meter off.
2. Apply more compound.
3. Tighten the pipe strap.
4. Restart the meter to see if warning clears.
B
1. Reposition transducers.
2. See if signal strength increases.
3. See if warning clears.
4. Restart the meter.
C
1. Relocate transducers.
2. See if signal strength increases.
3. See if warning clears.
4. Restart meter.
D
1. Turn meter off, and remove the transducers.
2. Grind the pipe surface if it is rough or coated.
3. Remount the transducers.
4. Restart the meter. See if warning clears.
General
Troubleshooting