If it should become necessary to calibrate the CO
2
controller, perform the
following procedure.
Start from the standard operating display (setpoint in bottom display,
actual CO
2
reading in the upper display).
1. Press the Advance key until “CAL” appears in the lower display.
2. Use the up and down arrows to make offset adjustment.
3. Press the Infinity key to exit the calibration menu.
4-10
Model 3940 Series
Thermo Electron Corporation
Section 4
Service
CO
2
Controller
Calibration
Figure 4-13.
Key Locations