42
Note
Basic CP mode configuration is complete. See page 47
(
2-2. Basic Configuration
) for more configuration options.
The current range and voltage range applies to all the
operating modes.
2-1-6. +CV Mode
Description
+CV mode can be added to CC, CR and CP mode.
The +CV settings apply to all applicable modes.
Operation
1.
Make sure the load is off.
2.
Press
Main
.
And select to Mode, I Range, and V Range.
3.
Set the +CV voltage level. (You may need to scroll
down to the +CV setting)
+CV:
OFF ~ rated 5%
Display
Mode
CP + CV
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
+CV
5.500
Timer2
0.025
Timer1
0.025
LOAD
RS232
01/Oct/2017
V
ms
ms
+CV setting
Note
The +CV settings apply to all the applicable operating
modes.
For example: The +CV settings made in CR mode will be
carried over to the +CV settings in CC and CP mode.
Note
Only in +CV settings, the external control is not possible.
See page 133 (
4-1-3. External Voltage Control
– Operation
)
for +CV settings with external control.