150
Syntax
SOURce[1|2]:VOLTage:UNIT {VPP|VRMS|DBM}
Example
SOUR1:VOLT:UNIT VPP
Sets the amplitude units to Vpp.
Query Syntax
SOURce[1|2]:VOLTage:UNIT?
Return
Parameter
VPP
Vpp
VRMS
Vrms
DBM
dBm
Example
SOUR1:VOLT:UNIT?
VPP
The amplitude units are set to Vpp.
8-8. Pulse Configuration Commands
The pulse chapter is used to control and output pulse waveforms. Unlike the
APPLy command, low level control is possible including setting the rise time, fall
time, period and pulse width.
Fall time
Rise time
Pulse Width
Period
50%
90%
10%
90%
50%
10%
8-8-1. SOURce[1|2]:PULSe:PERiod
Set
Query
Description
Sets or queries the pulse period. The default period is 1 ms.
Note
The pulse period must be greater than the pulse width and
edge time(1.6x) combined.
Pulse Width + (1.6 * Edge Time) < Period
If the edge time or pulse width are too great, they will
automatically be reduced to fit the period by the function
generator.
The PULSe:PERiod function will change the period for all