SBOA103C
Jan 2006
PGA309 Quick Start
System Reference Guide
77
Note that the
3 pressure 3 temperature
calibration algorithm
will compute values for nonlinearity error. This value needs
to be very small for this configuration because nonlinearity
correction is not used. This value will not be used to
generate the calibration tables in the EEPROM.
The value of Klin stored in the
EEPROM will be zero for this mode
because Vexc is disabled.
PGA309 Calibration
Spreadsheet, Sensor
Curvefit Tab