2
1
4
3
6
5
8
7
Introduction to PRBS Test
3
SLAA750 – July 2017
Copyright © 2017, Texas Instruments Incorporated
DAC38RF8x Test Modes
1.3
Hardware Setup
Follow these steps (see
) to set up the hardware:
Step 1.
Connect the TSW14J56 FMC interface connector (J4 of TSW14J56) to DAC38RF8x FMC
interface connector (J20 of DAC38RF8x EVM).
Step 2.
Connect a USB 2.0 Type A to Mini-B cable from the PC to DAC38RF8x EVM USB Mini-B
port (J16).
Step 3.
Connect a USB 3.0 Type A to Type B cable from the PC to TSW14J56 RevD USB 3.0 B port
(J9).
Step 4.
Connect a 5-V power supply to the DAC38RF8x EVM board using J21.
Step 5.
Connect a 5-V power supply to the TSW14J56 board using J11.
Step 6.
Turn on the TSW14J56 board by moving switch 6 to the ON position.
Step 7.
Connect the signal-generator output to LMKCLKIN (J4) of the DAC38RF8x EVM board.
•
Configure the signal generator to output a frequency of 368.64 MHz with an amplitude of
10 dBm.
•
Ensure that JP10 is removed from the board to enable internal clocking.
Step 8.
Attach an oscilloscope probe to the alarm pin of DAC38RF8x EVM board (TP9).
Figure 1. PRBS Hardware Setup