Installation
14
When using a cooler
The hot water can cause burns and damage wetted parts of
Testomat
®
808.
Water outlet
The feed water flows through the measuring chamber to the drain via
the outlet hose.
Remove the red plug from the outlet connection.
Connect the outlet connection of Testomat
®
808 to an outlet hose
(internal diameter 4 mm).
Feed the hose to an outlet.
Transportation plug!
The outlet is sealed with a plug to prevent leakage during transporta-
tion. Keep and store the plug for possible transportation at a later
date.
Connecting the power supply and devices
Risk of electric shocks during installation!
If the power supply is not disconnected prior to installation, it may
result in personal injuries, destruction of the product or damage to
plant parts.
Always disconnect the relevant plant parts before installing Tes-
tomat
®
808.
Only use tested cables with sufficient cross-sections for the con-
nections.
Risk of damages caused by electromagnetic fields!
If Testomat
®
808 or the connecting cables are installed parallel to
power cables or in close proximity to strong electromagnetic fields,
the instrument may be damaged or measurements incorrect.
Ensure that connecting cables are as short as possible.
Always install connecting cables and power cables separately.
Connect the instrument to the protective earth conductor (for
230/115 VAC).
Shield the instrument against strong electromagnetic fields.
CAUTION
!
WARNING
!
NOTE
NOTE