Error messages

Corona voltage out of range
The voltage in the unipolar corona charger is high or low. It does not necessarily mean that the instrument is 
not operating properly. If you see this warning, you need to check the corona voltage in the advanced menu:
If it is larger or equal to 5 kV, and remains so constantly, then it is appropriate to clean the corona wire. If it is 
between 2-3kV, this could be normal operation at low ambient pressure (i.e. you are at an alpine site or doing 
airborne measurements). If it is between 2-3kV and you are at or close to sea level, then it could be due to 
dirt on the counterelectrode of the charger.
In either case, you should check the display of the charging current, both with the charger on, and with the 
charger off. It should be 0nA with the charger off, and 10±1nA with the charger on. If it is not 0nA with the 
charger off, this means there is either dirt on the counterelectrode, or that you are operating the instrument 
at condensing conditions (too high relative humidity RH) and that a 

fi

 lm of water has formed on the countere-

lectrode insulation. If you think the RH could have been too high, keep the charger turned off, and continue 
to run the instrument in an environment with low relative humidity. The charging current should then go back 
to 0. If it does not, you have to clean the counterelectrode.
If the charging current is below 9nA with the charger on, and the corona voltage is >= 5kV, then the corona 
wire should be cleaned. Please concact the technical support and send the instrument for service.

Dirt on counterelectrode
This warning appears if the charging current is non zero when the corona charger is turned off. Follow the 
instructions in the section on corona voltage out of range above.

No or bad SD-card inserted
This warning appears if you try to start recording when no SD-card is inserted in the instrument, or if the 
SD-card is not formatted properly. If this warning appears despite that there is an SD-card in the instrument, 
then you should reformat the SD-Card.

File system error
If this error occurs, you should reformat your SD-card

Card full
This error occurs when your SD-card has no more space to write data. You should remove it, save all data 
on it, and reformat it.

General error
Contact instrument manufacturer. Email: [email protected]

The 

fi

 lter (or diffusion) stage current reached its maximal value (overrange)

At very high particle concentrations (larger than one million particles per cm3), the testo DiSCmini electro-
meter ampli

fi

 ers might reach their maximal level. In this case, all calculated signals (number, diameter, LDSA) 

are no longer correct. You can check where this occurred by pressing the 

fi

 lter or diffusion stage button on 

the toolbar in the software, to quickly see whether this was just a short episode, or whether the entire mea-
surement was not reliable.
Particle levels which cause the testo DiSCmini to reach its maximal electrometer currents should be avoided 
in any case – at such extreme particle levels, you will have to service your instrument very frequently! If you 
plan on doing measurements at high particle levels, you should use an external dilution system.

Your instrument requires a service due to excessive dirt in diffusion stage
The testo DiSCmini collects particles in both its measurement stages, which eventually become dirty and 
clog. The deposition characteristics in the diffusion stage will change with build-up of dirt, and the instrument 
readings become unreliable. To protect you from using an unreliable instrument, this warning appears when 
the integrated current deposited on the diffusion stage reaches a certain value , and we use it to recommend 
an instrument service. You can continue to use the instrument even if this warning appears, but you have 
to be aware that only the LDSA reading will then be reliable, whereas particle number and particle diameter 
are no longer reliable!

Summary of Contents for DiSCmini

Page 1: ...testo DiSCmini Diffusion Size Classifi er Miniature Instruction Manual Version V1 12 September 2016...

Page 2: ......

Page 3: ...icators and Connections 7 Aerosol Inlet 7 LCD screen 8 Main Screen 8 Raw Data Screen Main 8 Surface Area Screen Main 9 Advanced mode navigation 9 Charger Screen advanced 9 Flow Screen advanced 9 Heati...

Page 4: ...nd spare parts purchase ordering contact your local distributor or testo techsup port Email support nanoparticle testo de Sales support and customer service For commercial request please contact your...

Page 5: ...n not carried out by Testo SE Co KGaA or an authorised service partner use of not genuine spare parts Liability to content The content of this manual is generated with most accurateness Testo SE Co KG...

Page 6: ...Warnings Warnings for testo DiSCmini Caution Caution means be careful If do not follow the manual instruction you might result an in strument or accessories damage but no human injury Also Caution ref...

Page 7: ...y damage the instrument High particle concentrations To avoid premature aging of the instrument do not expose it to extremely high aerosol concentrations such as undiluted cigarette smoke Large partic...

Page 8: ...les will preferably deposit on it by diffusion and detected as an electrical current The remaining particles end up in a second stage the filter stage and are also detected as an electrical current Th...

Page 9: ...let The aerosol inlet is at the bottom of the instrument The impactor is used to remove larger particles 0 7 microns which would alter the measurements The impactor must always be fitted before starti...

Page 10: ...or heating and a blinking circle when recording On the lower right corner the remaining battery charge is displayed This is only an approximate value Pressing the right arrow button will lead you to t...

Page 11: ...advanced menu described below press 3 sec the menu button Press on menu to jump from one screen to the other Charger Screen advanced 3244V 9nA H OFF HIGHVOLT ON 3244V 9nA H OFF HIGHVOLT ON The charger...

Page 12: ...g table during the offset compensation Date Time Screen advanced SET DATE TIME 2000 04 18 29 SET DATE TIME 2000 04 18 29 You can set the clock and date in this screen Every element year month day will...

Page 13: ...s the record REC button The testo DiSCmini will indicate that it is recording by blinking a recording symbol in the upper right corner of its display Stop the recording by pressing the REC for few sec...

Page 14: ...een 2 and 4 hours depending on the current status of the battery The LED on the charger indicates if it is still charging or if charging is finished Measuring with testo DiSCmini The purpose of this a...

Page 15: ...ts pump once an hour for one minute and continues recording data testo DiSCmini does not use this data immediately instead the offline testo DiSCmini software detects the intervals where the pump was...

Page 16: ...ameter displays the average particle diameter over time in the graph panel LDSA displays the lung deposited surface area over time in the graph panel Induction correction Rapidly varying charge concen...

Page 17: ...Temperature displays the instrument temperature over time in the graph panel Flow displays the flow in the instrument over time in the graph panel Note that this is not actual flow in l min but the s...

Page 18: ...d in the instrument or if the SD card is not formatted properly If this warning appears despite that there is an SD card in the instrument then you should reformat the SD Card File system error If thi...

Page 19: ...act Testo or your local distri butor for service Noise levels too high Run testo DiSCmini with filtered and dried air for several hours to remove a water film on the insulators which can form at high...

Page 20: ...2E3 1E6 pt ccm 100nm 5E2 5E5 pt ccm Accuracy 30 in size and number typical Flow rate 1 0 L min 0 1 L min Operating conditions Pressure 800 1100 mbar abs ambient p max at inlet 20mbar Temperature 10 30...

Page 21: ......

Page 22: ...0970 3700 en 02 Testo SE Co KGaA Testo Stra e 1 79853 Lenzkirch Deutschland Tel 49 7653 681 5062 Fax 49 7653 681 95062 Web www testo particle com E Mail sales nanoparticle testo de...

Reviews: