30
On
was selected:
2
Use
/
to set the value for the upper alarm
threshold ( ) and confirm with
.
3
Use
/
to set the value for the lower alarm
threshold ( ) and confirm with
.
3
To rregister tthe rradio pprobe:
Radio probes may only be used in countries in which
they have been Type Approved (see application infor-
mation of the radio probe).
The setting function for radio probes is only available if
a radio module (accessory part) is inserted into the
measuring instrument.
º
See the chapter
C
OMMISSIONING
.
If no radio module is inserted:
º
Continue with objective
T
O SET
A
UTO
O
FF
.
Each radio probe has a probe ID (RF ID). This consists of
the last 3 digits of the serial no. and the position of the
slide switch in the radio probe (
H
or
L
).
Configuration mode is opened and
RF ID
and
Auto
are
lit.
The radio probe is switched on.
1
Select the desired option with
and confirm with
:
·
YES
: Switches automatic probe detection on
(recommended).
·
no
: Switches automatic probe detection off.
no
was selected:
2
Use
/
to set the probe ID manually and
confirm with
.
5. Operation