8
Maintenance
38
8.4
Firmware update
The current firmware version is available at www.testo.com.
There are two options available:
•
Performing an update using IRSoft or
•
Performing an update with the thermal imager
Downloading firmware
1
- Download firmware:
Firmware-testo-865-872.exe
2
- Unzip file: double-click on the exe. file
FW_T87x_Vx.xx.bin
is stored in the selected folder.
8.4.1
Performing an update using IRSoft
8.4.1.1
Preparing the camera
- The rechargeable battery is fully charged or the camera is connected
to the mains adapter.
1
- Connect the computer to the camera with the USB cable.
2
- Switch on the camera.
8.4.1.2
Performing an update
IRSoft is activated.
1
- Select
Camera
→
Configuration
.
2
- Click on
Configure thermal imager
.
The
Camera settings
window opens.
3
- Select
Instrument settings
→
Firmware Update
→
OK
.
FW_T87x_Vx.xx.bin
is displayed.
4
- Select
Open
.
Firmware update is performed.
Thermal imager is automatically restarted. Update complete.