5. Menu guide
18
( 0555.6743 only) supplement
available from versioin June 2007 on. For older versions an update
of the firmware is necessary
Reference value > -60°tpd
(rEF)
9
= +1
= Next digit
Carry out reset (rES)
nor = No reset
rS = Reset (return to
default setting))
Auswahl mit
Comments:
14
Can only be used in combination with a 2-point adjustment (cf.
chap 6.2)
15
In this setting point,
P1
and the relative humidity at the pressure
level
P1
are alternately displayed.
Please wait until the relative humidity has a constant value, and
only then press the SET button.
16
In this setting point,
P2
and the relative humidity at the pressure
level
P2
are alternately displayed.
Switch the switchover cock of the adjustment device and then wait
until the relative humidity has a constant value, and only then press
the SET button.
AdJ
rST
Return to
AdJ
Return to
rSt
Set pressure ratio
P
r
= P
1
/P
2
= +1
= Next digit
2PA
14
Return to
2PA
Return to
UNIT
Display value P
1
15
Display value P
2
16