Notes
25
Analyzing/testing the instrument:
The required instrument file is selected.
1
Click on
Test/analyze transmitter
button.
2
Perform tasks:
Options
· Carry out factory reset: Reset the parameter unit, scaling limits and adjustment to the factory settings
(Values instrument-specific, see specification plate).
· Transmitter tests (voltage supply via terminals required): Current reading displayed and analog output test
carried out for each channel; see below for procedure.
· Min./max. values: Change to display of minimum/maximum values.
Procedure: testing analog output (voltage supply via terminals required)
1 Choose channel.
2 Connect reference multimeter (min. 6.5 digits resolution) to the analog output terminals of the transmitter.
3 Set the default value in the P2A Software and click on
Activate
.
4 Compare value displayed with reference value of multimeter.
3
To leave the analyzing/test screen, click on
OK
.
Carrying out/resetting a 1-ppoint adjustment (offset):
A testo 400/600 with precision humidity probe is recommended as the
reference measuring instrument.
1
Click on
Adjust transmitter
button.
2
Select the channel under
1-point adjustment
.
3
Expose the reference measuring instrument and the instrument to be
adjusted to the same constant conditions and wait for equalization period
to lapse.
4
Enter reference value and perform adjustment by clicking on
Carry out 1-
point adjustment
.
To reset an offset value, click on
Set Offset to zero
.
5
Click on
OK
to confirm entries.
Carrying out a 2-ppoint adjustment:
1
Click on
Adjust transmitter
button.
2
Select the adjustment point under
2-point adjustment
.
3
Position the entire testo 6621-A01 (wall) or the probe shaft of the testo
6621-A02/A03 in a precision humidity generator/air conditioning cabinet
and route the USB cable to the P2A Software outwards.
4
To set the humidity generator: Temperature of 25 °C and humidity of
11.3 % RH or 75.3 % RH (depending on adjustment point chosen).
de
fr
es
it
pt
sv
nl
??
??
Summary of Contents for 6621