Tips and assistance
Questions and answers
Question
Possible causes/solutions
Hi
or
Lo
· Readings outside the measuring range (too high, too ·
low): Readings outside the measuring range (too
high, too low): for materials that have not been
specially dried or moistened, the measurement
results should always be within the measuring range.
·
Residual capacity <10 min: Change batteries.
Testing the instrument
function:
Test: ok
does
not light up
· Clean sensing electrodes and contacts on the top
of the protection cap.
· If the error message appears again, send the
instrument to Testo Customer Service.
Meas. electrodes defective/
worn
· Send instrument to Testo customer service.
The material‘s measured moisture content is relative to the dry
weight (0% moisture) of the relevant material. The stored material
characteristic curves were determined using the Darr (wet and dry
weighing) method. The specified measuring ranges are derived
from this.
Conversion to percent by weight
Percent by weight = (wet weight-dry weight) * 100 / dry weight
Example
Wet weight: 180 g
Dry weight: 150 g
Percent by weight: (180 - 150) * 100 / 150 = 20%
Conversion to water content
Water content % = (wet weight - dry weight) * 100 / wet weight
Example
Wet weight: 180 g
Dry weight: 150 g
Water content %: (180 - 150) * 100 / 180 = 16.6%
17
Summary of Contents for 606-1
Page 28: ...28...
Page 38: ...38...
Page 48: ...48...
Page 58: ...58...
Page 59: ...59 de en fr es it pt ru nl testo 606 1 C 2c OFF ON 10 Hold 2 59 testo 606 1...
Page 60: ...60 60 C Testo Testo...
Page 62: ...62 62 C 1 2 2x 1 5 AAA C 3 OFF ON 10 1 2 3...
Page 64: ...64 64 1 Test 2 C Test Test ok Test ok 3 Hold 1 2 2 x 1 5 AAA C 3...
Page 68: ...68...
Page 78: ......
Page 79: ......