3 Commissioning
14
The rib on the battery is in the position marking.
ATTENTION
Inserting the battery incorrectly will cause damage!
Ingress of moisture!
-
The rib must be positioned inside the position marking on the data
logger.
The service life of the batteries depends very much on the conditions
in the area of use. Under typical autoclave and freeze-drying
conditions (measuring cycle 10 s and service life 2 h/d), we
recommend replacing battery 0515 1900 after one year at the latest
and battery 0515 1901 after 50 days. Increased measuring cycles or
reduced measuring rates may result in a shorter service life.
Therefore, please pay attention to the battery level indicator in the PC
software.
3.2
Positioning the distance adapter in the
programming and readout unit
The data loggers are supplied with the appropriate distance adapters.
The testo 191 P1 data logger does not have a distance adapter and
can be placed in the programming and readout unit without using one.
1
Insert the distance adapter into the
socket first.
2
Push forward until the side brackets engage easily.