Calling up the function:
-
Measurements
-
OK
-
Finepress. probe
-
OK
-
E-Delta-P
-
OK
D.1.2.1 Individual measurement:
Performing the measurement:
1
single
-
OK
2 Start measuring:
Start
3 Stop measuring:
Stop
- The reading determined from
E-Delta-P
is recorded and copied into the Flue gas
menu, where it is displayed (if selected in the Display sequence menu).
D.1.2.2 Logger
Performing the measurement:
1
Logger
-
OK
2 Enter the measurement period:
Measurement period
-
Change
- enter value
(max. 1440 min)
3 Enter rate:
Rate
-
Change
- enter value (default value 1 sec)
The measuring rate must not be greater than the measurement period
4 Start measuring:
Start
- The residual time and the current reading are shown.
Options:
Print readings:
.
A printout of all readings is only possible in the “
Memory/location
” menu
(
Memory/location
-
OK
-
Data
-
Print all)
Display readings graphically:
Graph
.
From the graphic display back to the reading display:
Values
5 Stop measuring:
Stop
.
- The determined reading from
E-Delta-P
is displayed.
The logger program is always saved, even if it has not run through completely.
9