8 Maintaining the product
17
The instrument can become very warm during charging and should not
be held in your hand.
8.2
Cleaning the instrument
1
If the housing of the instrument is dirty, clean it with a damp cloth.
Do not use any aggressive cleaning agents or solvents! Mild household
cleaning agents and soap suds may be used.
Storage & transportation
To prevent contamination of the sensor, please do not store or transport the
instrument in an environment where any tobacco smoke, foul air, oils, greases,
silicones, evaporating liquids or gases are present. Any sensor that is
contaminated as a result of storage or transportation must be cleaned before
use, see Cleaning the sensor.
Regular inspection
Testo recommends having the gas leak detector inspected by an authorized
service centre every year.
8.3
Cleaning the sensor
Tobacco smoke, dirty air, oils, greases, silicones and evaporating liquids or
gases can leave deposits on the sensor surface. Possible consequences are
reduced sensitivity, distorted displays of gas concentration or display of a
background concentration. Clean the sensor if necessary
1
Switch on the instrument, allow it to initialize and then switch it off.
Repeat this procedure several times.
2
If it is dirty, clean the sensor head with a soft, dry cloth.
Switching on regularly
If the instrument is used infrequently, deposits may build up on the sensor.
Switching the instrument on prevents these deposits from building up on the
sensor. Testo recommends switching the instrument on regularly to avoid
deposits building up on the sensor.
Changing the sensor head
The instrument has a permanently installed sensor head that can only be
replaced by Testo Customer Service.