8 Maintaining the product 

17 

 

The instrument can become very warm during charging and should not 
be held in your hand. 

 

 

8.2 

Cleaning the instrument 

If the housing of the instrument is dirty, clean it with a damp cloth. 

 

 

Do not use any aggressive cleaning agents or solvents! Mild household 
cleaning agents and soap suds may be used. 

 

Storage & transportation

 

To prevent contamination of the sensor, please do not store or transport the 
instrument in an environment where any tobacco smoke, foul air, oils, greases, 
silicones, evaporating liquids or gases are present. Any sensor that is 
contaminated as a result of storage or transportation must be cleaned before 
use, see Cleaning the sensor. 

 

Regular inspection 

Testo recommends having the gas leak detector inspected by an authorized 
service centre every year. 

 

8.3 

Cleaning the sensor 

Tobacco smoke, dirty air, oils, greases, silicones and evaporating liquids or 
gases can leave deposits on the sensor surface. Possible consequences are 
reduced sensitivity, distorted displays of gas concentration or display of a 
background concentration. Clean the sensor if necessary 

 

Switch on the instrument, allow it to initialize and then switch it off. 

Repeat this procedure several times. 

 

If it is dirty, clean the sensor head with a soft, dry cloth. 

 

Switching on regularly 

If the instrument is used infrequently, deposits may build up on the sensor. 
Switching the instrument on prevents these deposits from building up on the 
sensor. Testo recommends switching the instrument on regularly to avoid 
deposits building up on the sensor. 

 

 

Changing the sensor head 

The instrument has a permanently installed sensor head that can only be 
replaced by Testo Customer Service. 

Summary of Contents for 0560 3163

Page 1: ...testo 316 1 EX gas leak detector 0560 3163 Instruction manual...

Page 2: ......

Page 3: ...tion 8 6 2 1 Establishing a Bluetooth connection to the testo Smart App 8 7 Using the product 9 7 1 Controls 9 7 1 1 Implementing settings 11 7 2 Carrying out a function test 13 7 3 Carrying out gas d...

Page 4: ......

Page 5: ...Do not operate the instrument if there are signs of damage on the housing mains unit or connected cables Dangers may also arise from objects to be measured or the measuring environment Always comply...

Page 6: ...tteries in accordance with the valid legal specifications At the end of its useful life deliver the product to the separate collection point for electric and electronic devices observe local regulatio...

Page 7: ...lammable substance in the air has a lower explosive limit LEL and an upper explosive limit UEL The air gas mixture is flammable anywhere between these two limits potentially leading to an explosion cr...

Page 8: ...play 5 Control keys 6 Charging socket for rechargeable battery Symbol explanation Do not allow children under 6 years of age to play with batteries Do not throw batteries in the trash Do not charge ba...

Page 9: ...on the longer this additional warming up phase should be Please note that the instrument switches itself off by default after 10 min of inactivity 1 Press and hold down 1 sec the On Off key Warm up ph...

Page 10: ...ing is confirmed by AOFF ON or AOFF OFF 6 2 Establishing a Bluetooth connection The instrument can be connected to the testo Smart App via Bluetooth connection The gas leak detector is switched on 6 2...

Page 11: ...The app is installed on the smartphone and connected to the instrument via Bluetooth Settings and controls are primarily carried out on the instrument and are transferred to the app To a lesser extent...

Page 12: ...measurement data report Searching Favourite Deleting Further information Display report Multiple selection The Main menu can be accessed via the icon at top left To exit the main menu select a menu o...

Page 13: ...ng and holding the key 1 sec Adjustable functions Ensure correct settings all settings are transferred immediately There is no Cancel function Function Setting options comments On Off long press Switc...

Page 14: ...of the alarm sound increases as the concentration increases Unit long press Switch between PPM and LEL Measuring mode MODE 1 for gas detection MODE 2 for locating leaks Max zeroing long press Zero max...

Page 15: ...rer Testing natural gas lines or hydrogen lines Methane main component of natural gas or hydrogen are lighter than air detection should be carried out above the pipe suspected leak Testing propane and...

Page 16: ...falling concentration values are output At the beginning two bar segments are displayed in the middle At the same time the sensor LED flashes in a colour corresponding to the concentration value and a...

Page 17: ...k it is advisable in the case of these larger leaks to switch to MODE 2 in order to receive feedback via audio and flashing signals If the warning threshold 100 ppm is exceeded the sensor LED and disp...

Page 18: ...using max 0 After the measurement 1 Ventilate the sensor thoroughly after each use To do this place the instrument in fresh air for approx 2 minutes before using it again 8 Maintaining the product 8...

Page 19: ...r inspected by an authorized service centre every year 8 3 Cleaning the sensor Tobacco smoke dirty air oils greases silicones and evaporating liquids or gases can leave deposits on the sensor surface...

Page 20: ...se time 2 sec Leak alarm 3 colour LED on sensor head 3 colour display backlighting Acoustic App Operating temperature 5 to 50 C Operating humidity 0 to 80 RH Operating altitude 2000 m Storage temperat...

Page 21: ...ensor error The error code is shown on the display 10 2 1 Hard Reset 1 If you encounter any problems with the firmware press and hold the ON OFF key for a long time 4 sec to perform a reset 10 3 Acces...

Page 22: ...Testo SE Co KGaA Celsiusstr 1 79822 Titisee Neustadt Germany Phone 49 0 7653 681 0 Email info testo de www testo com 0970 3263 en 01 08 2022...

Reviews: