Chapter 7 – Maintenance
Page 7-8
TestEquity TEC1 Thermoelectric Temperature Chamber
EZ-Zone Setup Menu Parameters
Setup Page access from Home Page:
Press both the Up
¿
and Down
¯
keys for six seconds.
[``Ai]
will appear in the upper display and
[`Set]
will appear in the lower display.
Setup Menu\Analog Input Menu\Input1
Parameter
Prompt
Value
Sensor Type
SEn
r0.1H (
RTD 100 Ω
)
RTD Leads
rt.L
3
Filter Time
FiL
0.5
Error Latching
i.ER
oFF
Decimal
dEC
0.0
Setup Menu\Loop Menu
Parameter
Prompt
Value
Heat Algorithm
h.Ag
Pid
Cool Algorithm
C.Ag
Pid
Cool Output Curve
C.Cr
off
Tru-Tune Enable
t.tUn
no
Autotune Aggressiveness
t.AGr
Crit
Peltier Delay
P.dL
0.2
User Failure Action
UFa
oFF
Input Error Failure
FAiL
oFF
Manual Power
MAn
0.0
Open Loop Detect Enable
L.dE
no
Ramp Action
rP
oFF
Low Set Point
L.SP
9.0 °C
High Set Point
h.SP
61.0 °C
Set Point Low Open Loop
SP.Lo
-100.0 %
Set Point High Open Loop
SP.hi
100.0 %
Setup Menu\Output Menu\Output 1
Parameter
Prompt
Value
Function
Fn
hEAt
Function Instance
Fi
1
Control
o.Ct
Ftb
Time Base
o.tb
0.1
Low Power Scale
o.Lo
0.0 %
High Power Scale
o.hi
100.0 %
Setup Menu\Output Menu\Output 2
Parameter
Prompt
Value
Function
Fn
CooL
Function Instance
Fi
1
Control
o.Ct
Ftb
Time Base
o.tb
0.1
Low Power Scale
o.Lo
0.0
High Power Scale
o.hi
100.0