Chapter 4 – Temperature Controller
Page 4-4
TestEquity TEC1 Thermoelectric Temperature Chamber
Profile Programming
The EZ-Zone Temperature Controller can be programmed to store up to 40 total steps into as
many as 4 profiles. You do not need a computer to enter a profile – it can be easily done through
the controller’s front panel keys. A Profile is a set of instructions programmed as a sequence of
steps. The controller handles the profile steps automatically, in sequence.
Refer to Chapter 7 of the “EZ-Zone PM User’s Manual” for detailed instructions on using the
Profile function.
To enter the Profiling Page from Home Page:
Press the Advance Key
‰
for three seconds
until
[ProF]
appears in the lower display.
NOTES:
TestEquity recommends having the End step type to be Hold or User. TestEquity does NOT
recommend using an end step type of Off. This does not turn off the chamber fans and the
temperature will be uncontrolled.
Input events and output events are not included in the controller supplied with the TEC1
chamber. Ignore any mention of them in the “EZ-Zone PM User’s Manual”. Only the capabilities
of a model PM6R1CC-2AAAAAA, as described on page 112 of the “EZ-Zone PM User’s
Manual” are applicable.