Rev. 01/29/10 MKB  

ACL 350 

 Page 2 of 9 

ACL 350 DIGITAL  STATIC LOCATOR 

 

 

Before using this, or any other electronic piece of equipment, carefully read the instruction 
manual completely.  
 
I.   

BEFORE USING THE INSTRUMENT Examine the instrument for damage, 
contamination (excessive dirt, grease, etc.) and defects.  

 

If any abnormal conditions exist, do not attempt to make any measurements.  

 
 
II.   

SAFETY  

 

A.  The following safety information must be observed to insure maximum personal 

safety during the operation of this meter.  

1. 

DO NOT use the meter if the meter looks damaged, or if you suspect that 

the meter is not operating properly.  

2. 

DO NOT touch the front sensing head while operating this meter.  

3. 

Measuring voltage which exceeds the limits of this meter may damage 
the meter and expose the operator to a shock hazard. Always recognize 
the meter voltage limits as stated.  

 
 
III.   

FUNCTION  

 

The ACL 350 Static Locator is a non-contact meter 
that indicates electrostatic field voltage and polarity on 
charged surfaces. This easy to use instrument comes 
equipped with a Power On/Hold button, a zero button 
and a unique retractable one inch spacer. The meter 
automatically shuts off after approximately 90 
seconds. The unit's grounding path is through its 
conductive case, so there is often no need to wear a 
grounded wrist strap during normal operations. The 
easy to read 3-1/2 digit LCD display automatically 
updates three times per second. Measurements are 
dependent on the distance from the object being 
measured.  

 
 
IV.   

ACCURACY  

 

A.  The accuracy of the ACL 350 is dependent upon three factors.  

1. 

The instrument must be properly grounded and zeroed.  

2. 

The distance from the front edge of the case to the target or surface 
under examination must be accurate and stable.  

 

Summary of Contents for ACL 350

Page 1: ...OPERATIONS MANUAL ACL 350 DIGITAL STATIC LOCATOR...

Page 2: ...CY ACL 350 Page 2 BATTERY INSTALLATION MAINTENANCE ACL 350 Page 3 Battery Replacement ACL 350 Page 3 Cleaning ACL 350 Page 3 STANDARD OPERATION ACL 350 Page 4 BASIC OPERATION WHILE USING THE ONE INCH...

Page 3: ...se the operator to a shock hazard Always recognize the meter voltage limits as stated III FUNCTION The ACL 350 Static Locator is a non contact meter that indicates electrostatic field voltage and pola...

Page 4: ...Remove the battery from its compartment and disconnect it from the power terminals 6 Connect a new battery to the power terminals and install it in the battery compartment 7 Replace the battery cover...

Page 5: ...ick up noise from fluorescent lights electronics and air movement D Point the sensor plate toward the target and move to a point one inch from the surface to be measured Thus there should be a one inc...

Page 6: ...case beneath the display and operating buttons 2 The case of the instrument is conductive and is the reference for measurement B Press the POWER ON HOLD button and release C Face the static locator a...

Page 7: ...l it rests parallel to the case with its point toward the battery compartment VIII OPERATION WHILE USING THE ONE INCH SPACER AS A DISTANCE REFERENCE In some cases you may wish to use the spacer to est...

Page 8: ...urement target surface At this point there is approximately one inch space between the front sensing end of the case and the target 1 Note the meter reading 2 To hold the reading press and hold the PW...

Page 9: ...when in hold mode Spacer Retractable 1 Inch spacer Measurement Ranges 1 inch 0 20 000 volts in 100 volt increments 1 10 inch 0 2 000 volts in 10 volt increments Extended Ranges 4 inches 40 000 volts...

Page 10: ...0 1 000v 20 2 000v 1 0 1 100 v 1 1000v 10 10 000v 20 20 000v 4 0 1 2 x 100 v 200 v 1 2 x 1000v 2000v 10 2 x 10 000v 20 000v 20 2 x 20 000v 40 000v 6 5 0 1 3 x 100 v 300 v 1 3 x 1000v 3000v 10 3 x 10 0...

Reviews: