Chapter 3 – Operation
TestEquity 123C Temperature Chamber
Page 3-3
Loading the Chamber
!
WARNING:
Do NOT put items in the chamber that could burn or explode at high
temperatures. This chamber uses open wire heating elements which generate
surface temperatures over 1000ºF. This is NOT an explosion-proof chamber.
!
WARNING:
Do NOT put items in the chamber which can emit corrosive vapors or
substances.
!
WARNING:
This chamber is NOT a curing oven. There are NO provisions for venting
fumes.
!
WARNING:
The chamber door must remain closed while the chamber is operating. If you
need to open the door while the chamber is operating, wear safety goggles to
prevent the high velocity airflow from blowing particles or objects into your
eyes.
!
WARNING:
This chamber operates at extreme temperatures. Avoid contact with air,
objects, and surfaces that are hot or cold to prevent severe burns or frostbite.
Protective gloves are recommended.
!
CAUTION:
If your test sample is energized, it may be capable of raising the workspace
temperature beyond safe limits. This could occur if your test sample exceeds
the live load rating of the chamber or if the chamber’s refrigeration system
fails. This chamber has a set of contacts that can be used to remove power to
your test sample if the Limit Controller’s temperature limits are exceeded.
!
CAUTION:
To prevent damage to your test sample and the chamber’s compressors, do not
exceed the live load rating of the chamber.
Live Load Capacity for Model 123C
Temp +23°C
0°C
–40°C –55°C –65°C
Watts
500 W 400 W 225 W 175 W 100 W
Port Plugs
Foam port plugs are provided with a gray silicone surface on one side. The port plug must be
inserted with the gray silicone surface facing the inside of the chamber. Port plugs should be
considered expendable and be replaced when they no longer provide a good seal.